Commands by Subsystem 5
Agilent InfiniiVision 6000 Series Oscilloscopes Programmer's Guide 335
:MEASure:FREQuency
(see page 798)
Command Syntax
:MEASure:FREQuency [<source>]
<source> ::= {<digital channels> | CHANnel<n> | FUNCtion | MATH}
<digital channels> ::= DIGital0,..,DIGital15 for the MSO models
<n>::={1|2|3|4}forthefour channel oscilloscope models
<n> ::= {1 | 2} for the two channel oscilloscope models
The :MEASure:FREQuency command installs a screen measurement and
starts a frequency measurement. If the optional source parameter is
specified, the current source is modified.
IF the edge on the screen closest to the trigger reference is rising:
THEN frequency = 1/(time at trailing rising edge - time at leading rising
edge)
ELSE frequency = 1/(time at trailing falling edge - time at leading falling
edge)
Query Syntax
:MEASure:FREQuency? [<source>]
The :MEASure:FREQuency? query measures and outputs the frequency of
the cycle on the screen closest to the trigger reference.
Return Format
<source><NL>
<source> ::= frequency in Hertz in NR3 format
See Also "Introduction to :MEASure Commands" on page 324
":MEASure:SOURce" on page 349
":MEASure:PERiod" on page 339
Example Code "Example Code" on page 350
NOTE
This command is not available if the source is FFT (Fast Fourier Transform).