Agilent Technologies InfiniiVision 6000 Series Manual de Serviço Página 510

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 934
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 509
510 Agilent InfiniiVision 6000 Series Oscilloscopes Programmer's Guide
5 Commands by Subsystem
:TRIGger:EBURst:SLOPe
(see page 798)
Command Syntax
:TRIGger:EBURst:SLOPe <slope>
<slope> ::= {NEGative | POSitive}
The :TRIGger:EBURst:SLOPe command specifies whether the rising edge
(POSitive) or falling edge (NEGative) of the Nth edge in a burst will
generate a trigger.
Query Syntax
:TRIGger:EBURst:SLOPe?
The :TRIGger:EBURst:SLOPe? query returns the current Nth edge in a
burst slope.
Return Format
<slope><NL>
<slope> ::= {NEG | POS}
See Also "Introduction to :TRIGger Commands" on page 479
":TRIGger:EBURst:IDLE" on page 509
":TRIGger:EBURst:COUNt" on page 508
Vista de página 509
1 2 ... 505 506 507 508 509 510 511 512 513 514 515 ... 933 934

Comentários a estes Manuais

Sem comentários