Agilent Technologies InfiniiVision 6000 Series Manual de Serviço Página 232

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 934
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 231
232 Agilent InfiniiVision 6000 Series Oscilloscopes Programmer's Guide
5 Commands by Subsystem
:CHANnel<n>:BWLimit
(see page 798)
Command Syntax
:CHANnel<n>:BWLimit <bwlimit>
<bwlimit> ::= {{1 | ON} | {0 | OFF}
<n>::={1|2|3|4}forthefour channel oscilloscope models
<n> ::= {1 | 2} for the two channel oscilloscope models
The :CHANnel<n>:BWLimit command controls an internal low-pass filter.
When the filter is on, the bandwidth of the specified channel is limited to
approximately 25 MHz.
Query Syntax
:CHANnel<n>:BWLimit?
The :CHANnel<n>:BWLimit? query returns the current setting of the
low- pass filter.
Return Format
<bwlimit><NL>
<bwlimit> ::= {1 | 0}
See Also "Introduction to :CHANnel<n> Commands" on page 230
Vista de página 231