712 Agilent InfiniiVision 6000 Series Oscilloscopes Programmer's Guide
7 Obsolete and Discontinued Commands
:DISPlay:ORDer
(see page 798)
Query Syntax
:DISPlay:ORDer?
The :DISPlay:ORDer? query returns a list of digital channel numbers in
screen order, from top to bottom, separated by commas. Busing is
displayed as digital channels with no separator. For example, in the
following list, the bus consists of digital channels 4 and 5: DIG1, DIG4
DIG5, DIG7.
Return Format
<order><NL>
<order> ::= Unquoted ASCII string
See Also ":DIGital<n>:POSition" on page 253
Example Code
' DISP_ORDER - Set the order the channels are displayed on the
' analyzer. You can enter between 1 and 32 channels at one time.
' If you leave out channels, they will not be displayed.
' Display ONLY channel 0 and channel 10 in that order.
myScope.WriteString ":DISPLAY:ORDER 0,10"
Example program from the start: "VISA COM Example in Visual Basic" on
page 824
NOTE
The :DISPlay:ORDer command is an obsolete command provided for compatibility to
previous oscilloscopes. This command is only available on the MSO models.
NOTE
A return value is included for each digital channel. A return value of NONE indicates that a
channel is turned off.