Agilent Technologies 8935 Series Manual do Utilizador Página 96

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 352
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 95
96 Chapter 1
GPIB Command Dictionary
AFG2|ENCoder subsystem
:NAMPs|NTACs:FVC:RATE <real number>
:NAMPs|NTACs:FVC:RATE?
These commands set/query the data rate of the encoded message.
This command utilizes the “Real Number Setting Syntax” on page 299
but does not use the :STATe command.
Syntax
AFG2:NAMP:FVC:RATE <real number>
AFG2:NAMP:FVC:RATE?
Screen/field equivalent
AFG2:NAMP:FVC:RATE controls the Data Rate field on the
SIGNALING ENCODER screen when the Mode field is set to
NAMP-NTAC, and the Channel field is set to Voice.
:NAMPs|NTACs:FVC:AM <real number>
:NAMPs|NTACs:FVC:AM?
This command sets/queries the data level when the AFG2:DEST ‘AM’
command is used. The data level units are %.
This command utilizes the “Real Number Setting Syntax” on page 299.
Syntax
AFG2:NAMP:FVC:AM
AFG2:NAMP:FVC:AM?
Screen/field equivalent
AFG2:NAMP:FVC:AM controls the Data Level field when AFGen2 To
field is set to AM on the SIGNALING ENCODER screen when the Mode
field is set to NAMP-NTAC, and the Channel field is set to Voice.
Vista de página 95
1 2 ... 91 92 93 94 95 96 97 98 99 100 101 ... 351 352

Comentários a estes Manuais

Sem comentários