Chapter 1 111
GPIB Command Dictionary
CANalyzer subsystem
:POWer:ZERO
This command zeroes the power meter for an average power
measurement.
Syntax
CAN:POW:ZERO
Screen/field equivalent
CAN:POW:ZERO controls the Pwr Zero field on the CDMA ANALYZER
screen. This field is only available when the average power
measurement is selected.
:POWer:ZERO:MODE <string>
:POWer:ZERO:MODE?
These commands set/query whether the power meter is automatically
zeroed or if the zero function must be manually activated. This
command is used for average power measurements.
Syntax
CAN:POW:ZERO:MODE ‘Auto’
‘Manual’
CAN:POW:ZERO:MODE?
Screen/field equivalent
CAN:POW:ZERO:MODE controls the Auto Zero field on the CDMA
ANALYZER screen. This field is only available when the average power
measurement is selected.