Página 1 - Oscilloscopes

 Agilent 1000 Series OscilloscopesUser’s Guide

Página 2

10 Agilent 1000 Series Oscilloscopes User’s Guide Contents5 Saving, Recalling, and Printing Data 113Saving and Recalling Data 114To save and recall wa

Página 3

4 Making Measurements100 Agilent 1000 Series Oscilloscopes User’s GuideVoltage MeasurementsThere are 10 automatic voltage measurements:• Vmax (Maximu

Página 4 - In This Book

Making Measurements 4Agilent 1000 Series Oscilloscopes User’s Guide 101Vmin (Minimum Voltage)The minimum amplitude. The most negative peak voltage me

Página 5

4 Making Measurements102 Agilent 1000 Series Oscilloscopes User’s GuideVrms (Root-Mean-Square Voltage)The true root- mean- square voltage over the en

Página 6

Making Measurements 4Agilent 1000 Series Oscilloscopes User’s Guide 103Time MeasurementsThere are 12 automatic time measurements plus the hardware fr

Página 7

4 Making Measurements104 Agilent 1000 Series Oscilloscopes User’s GuideFrequencyMeasures the frequency of a waveform. See Figure 41 on page 103.Rise

Página 8

Making Measurements 4Agilent 1000 Series Oscilloscopes User’s Guide 105Positive Pulse WidthMeasures the positive pulse width of a waveform.Negative P

Página 9

4 Making Measurements106 Agilent 1000 Series Oscilloscopes User’s GuideDelay Between Rising EdgesMeasures the delay between two waveforms using the r

Página 10 - Contents

Making Measurements 4Agilent 1000 Series Oscilloscopes User’s Guide 107Phase Between Rising EdgesMeasures the phase between two waveforms using the r

Página 11

4 Making Measurements108 Agilent 1000 Series Oscilloscopes User’s GuideCounter (Frequency)The 1000 Series oscilloscopes have an integrated 6- digit h

Página 12 - Index 153

Making Measurements 4Agilent 1000 Series Oscilloscopes User’s Guide 109Making Cursor MeasurementsYou can use the [Cursors] front panel key to select

Página 13

ContentsAgilent 1000 Series Oscilloscopes User’s Guide 11 Performing Mask Tests 131To enable/disable mask tests 131To select the source channel for ma

Página 14

4 Making Measurements110 Agilent 1000 Series Oscilloscopes User’s GuideTo use manually adjustable cursorsYou can set up two parallel, manually adjust

Página 15

Making Measurements 4Agilent 1000 Series Oscilloscopes User’s Guide 111To use tracking cross-hair cursorsYou can set up one or two manually adjustabl

Página 16

4 Making Measurements112 Agilent 1000 Series Oscilloscopes User’s GuideTo display cursors for automatic measurements1 Press [Cursors].2 In the Curso

Página 17 - Getting Started

113Agilent 1000 Series OscilloscopesUser’s Guide 5Saving, Recalling, and Printing DataSaving and Recalling Data 114Using the Disk Manager 118Printi

Página 18

5 Saving, Recalling, and Printing Data114 Agilent 1000 Series Oscilloscopes User’s GuideSaving and Recalling DataUsing the oscilloscope’s [Save/Recal

Página 19

Saving, Recalling, and Printing Data 5Agilent 1000 Series Oscilloscopes User’s Guide 1153 Continue pressing the Storage softkey or turn the entry k

Página 20

5 Saving, Recalling, and Printing Data116 Agilent 1000 Series Oscilloscopes User’s GuideTo save to or recall from internal storage:a Press Internal.

Página 21 - Step 4. Input a waveform

Saving, Recalling, and Printing Data 5Agilent 1000 Series Oscilloscopes User’s Guide 1174 To specify whether oscilloscope parameters be saved along w

Página 22 - Step 5. Use Auto-Scale

5 Saving, Recalling, and Printing Data118 Agilent 1000 Series Oscilloscopes User’s GuideUsing the Disk ManagerWhen a USB drive is connected to the fr

Página 23 - Getting Started 1

Saving, Recalling, and Printing Data 5Agilent 1000 Series Oscilloscopes User’s Guide 119To switch between files, path, and directory panes1 In the Di

Página 24 - Step 6. Compensate probes

12 Agilent 1000 Series Oscilloscopes User’s Guide ContentsIndex 153

Página 25 - High Frequency Compensation

5 Saving, Recalling, and Printing Data120 Agilent 1000 Series Oscilloscopes User’s GuideTo edit folder/file namesIn the folder/file name edit dialog:

Página 26 - DSO1024A

Saving, Recalling, and Printing Data 5Agilent 1000 Series Oscilloscopes User’s Guide 121To delete foldersIn the directory pane (see “To switch betwee

Página 27

5 Saving, Recalling, and Printing Data122 Agilent 1000 Series Oscilloscopes User’s GuideTo recall filesIn the files pane (see “To switch between file

Página 28

Saving, Recalling, and Printing Data 5Agilent 1000 Series Oscilloscopes User’s Guide 123Printing ScreensYou can print oscilloscope display screens to

Página 29

5 Saving, Recalling, and Printing Data124 Agilent 1000 Series Oscilloscopes User’s GuideTo choose a PictBridge printerYou can print to a PictBridge c

Página 30 - Figure 8 Oscilloscope Display

Saving, Recalling, and Printing Data 5Agilent 1000 Series Oscilloscopes User’s Guide 125To print with inverted screen colors1 Press [Print].2 In the

Página 31

5 Saving, Recalling, and Printing Data126 Agilent 1000 Series Oscilloscopes User’s GuideTo copy a screen to the printer1 Press [Print].2 In the Prin

Página 32

127Agilent 1000 Series OscilloscopesUser’s Guide 6Oscilloscope Utility SettingsDisplaying System Information 128Turning Sound ON or OFF 128Setting

Página 33 - Hole for Kensington lock

6 Oscilloscope Utility Settings128 Agilent 1000 Series Oscilloscopes User’s GuideDisplaying System InformationTo display the oscilloscope’s system in

Página 34

Oscilloscope Utility Settings 6Agilent 1000 Series Oscilloscopes User’s Guide 129Setting and Displaying the Date and TimeTo set and display the oscil

Página 35 - Displaying Data

Agilent 1000 Series Oscilloscopes User’s Guide 13 FiguresFigure 1. Power Switch 19Figure 2. [Default Setup] Key 20Figure 3. [Auto-Scale] Key 22Figure

Página 36 - Position knob

6 Oscilloscope Utility Settings130 Agilent 1000 Series Oscilloscopes User’s GuideSetting the Language (Menu and Help)To set the language used in menu

Página 37

Oscilloscope Utility Settings 6Agilent 1000 Series Oscilloscopes User’s Guide 131Performing Mask TestsThe mask test function monitors waveform change

Página 38 - Slow Scan Mode

6 Oscilloscope Utility Settings132 Agilent 1000 Series Oscilloscopes User’s GuideTo run/stop a mask test1 In the Mask Test menu ([Utility]>Mask

Página 39

Oscilloscope Utility Settings 6Agilent 1000 Series Oscilloscopes User’s Guide 133To set the mask test output condition1 In the Mask Test menu ([Utili

Página 40 - Zoomed window view

6 Oscilloscope Utility Settings134 Agilent 1000 Series Oscilloscopes User’s GuideTo adjust a mask’s horizontal failure margin1 In the Mask menu ([Uti

Página 41 - X-Y Format

Oscilloscope Utility Settings 6Agilent 1000 Series Oscilloscopes User’s Guide 135To select internal/external mask storage location1 In the Mask menu

Página 42 - To view the sample rate

6 Oscilloscope Utility Settings136 Agilent 1000 Series Oscilloscopes User’s GuideSetting PreferencesThe oscilloscope’s Preference menu lets you set s

Página 43 - Math REF

Oscilloscope Utility Settings 6Agilent 1000 Series Oscilloscopes User’s Guide 137To select the USB device port functionThe (square) USB device port o

Página 44 - To adjust the vertical scale

6 Oscilloscope Utility Settings138 Agilent 1000 Series Oscilloscopes User’s GuideRunning Self-CalibrationThe automatic calibration routine adjusts th

Página 45 - To specify channel coupling

139Agilent 1000 Series OscilloscopesUser’s Guide 7Specifications and CharacteristicsEnvironmental Conditions 140Measurement Category 141Specificati

Página 46 - AC Coupling Status

14 Agilent 1000 Series Oscilloscopes User’s Guide FiguresFigure 28. Sample Rate and Oscilloscope Bandwidth 69Figure 29. Real-Time Sampling Mode 71Figu

Página 47 - Bandwidth limit OFF

7 Specifications and Characteristics140 Agilent 1000 Series Oscilloscopes User’s GuideEnvironmental ConditionsOvervoltage CategoryThis product is int

Página 48 - Bandwidth ON Status

Specifications and Characteristics 7Agilent 1000 Series Oscilloscopes User’s Guide 141Measurement CategoryThe 1000 Series oscilloscope is intended to

Página 49 - To use a digital filter

7 Specifications and Characteristics142 Agilent 1000 Series Oscilloscopes User’s GuideSpecificationsNOTEAll specifications are warranted. Specificati

Página 50 - To invert a waveform

Specifications and Characteristics 7Agilent 1000 Series Oscilloscopes User’s Guide 143CharacteristicsNOTEAll characteristics are the typical performa

Página 51 - Displaying Data 2

7 Specifications and Characteristics144 Agilent 1000 Series Oscilloscopes User’s GuideAC coupled1: DSO102xA: 5 Hz to 200 MHzDSO101xA: 5 Hz to 100 MHz

Página 52 - To specify the channel units

Specifications and Characteristics 7Agilent 1000 Series Oscilloscopes User’s Guide 145Table 8 Horizontal System CharacteristicsName Typical ValueRang

Página 53 - Math Scale

7 Specifications and Characteristics146 Agilent 1000 Series Oscilloscopes User’s GuideTable 9 Trigger System CharacteristicsName Typical ValueSources

Página 54

Specifications and Characteristics 7Agilent 1000 Series Oscilloscopes User’s Guide 147Table 11 Measurement FeaturesName Typical ValueAutomatic measur

Página 55

7 Specifications and Characteristics148 Agilent 1000 Series Oscilloscopes User’s GuideTable 13 StorageName Typical ValueSave/Recall internal: 10 setu

Página 56 - FFT Resolution

Specifications and Characteristics 7Agilent 1000 Series Oscilloscopes User’s Guide 149Table 15 General CharacteristicsName Typical ValuePhysical size

Página 57 - To save a reference waveform

Agilent 1000 Series Oscilloscopes User’s Guide 15 TablesTable 1. Agilent 1000 Series Oscilloscope Models 3Table 2. Auto-Scale Default Settings 23Table

Página 58

7 Specifications and Characteristics150 Agilent 1000 Series Oscilloscopes User’s GuideThe N2739A rack mount kit is available for placing a 1000 Serie

Página 59 - Changing the Display Settings

151Agilent 1000 Series OscilloscopesUser’s Guide ASafety NoticesWarnings 151Safety Symbols 152This apparatus has been designed and tested in

Página 60 - To set waveform persistence

ASafety Notices 152 Agilent 1000 Series Oscilloscopes User’s Guide• Whenever it is likely that the ground protection is impaired, you must make the

Página 61 - To change the grid

Agilent 1000 Series Oscilloscopes User’s Guide 153 IndexNumerics50% trigger level, 83AAC channel coupling, 45AC coupling, 54AC trigger coupling, 87, 9

Página 62 - To invert screen colors

154 Agilent 1000 Series Oscilloscopes User’s GuideIndex Fall Time measurement, 104FFT (Fast Fourier Transform) math function, 53, 54FFT resolution, 56

Página 63 - To select screen persistence

Agilent 1000 Series Oscilloscopes User’s Guide 155Index nonvolatile memory locations, 113Normal acquisition mode, 73normal adjustment, 44NTSC standard

Página 64 - 2 Displaying Data

156 Agilent 1000 Series Oscilloscopes User’s GuideIndex system information, displaying, 128Ttemperature at last calibration, 142theory, sampling, 66ti

Página 66 - Aliasing

Agilent Technologies, Inc. 2008-2009Printed in Malaysia 10/09Second Edition, October 2009*54130-97014*54130-97014www.agilent.com

Página 67

16 Agilent 1000 Series Oscilloscopes User’s Guide Tables

Página 68

17Agilent 1000 Series OscilloscopesUser’s Guide 1Getting StartedStep 1. Inspect the package contents 18Step 2. Turn on the oscilloscope 19Step 3. L

Página 69 - Oscilloscope Rise Time

1Getting Started18 Agilent 1000 Series Oscilloscopes User’s GuideStep 1. Inspect the package contents1 Inspect the shipping container for damage.Keep

Página 70

Getting Started 1Agilent 1000 Series Oscilloscopes User’s Guide 19Step 2. Turn on the oscilloscopeThe next few steps (turning on the oscilloscope, lo

Página 71 - Real-Time Sampling

Agilent 1000 Series Oscilloscopes User’s GuideNotices© Agilent Technologies, Inc. 2008-2009No part of this manual may be reproduced in any form or by

Página 72 - Memory Depth and Sample Rate

1Getting Started20 Agilent 1000 Series Oscilloscopes User’s GuideStep 3. Load the default oscilloscope setupYou can recall the factory default setup

Página 73 - Choosing the Acquisition Mode

Getting Started 1Agilent 1000 Series Oscilloscopes User’s Guide 21Step 4. Input a waveform1 Input a waveform to a channel of the oscilloscope.Use one

Página 74

1Getting Started22 Agilent 1000 Series Oscilloscopes User’s GuideStep 5. Use Auto-ScaleThe oscilloscope has an auto- scale feature that automatically

Página 75

Getting Started 1Agilent 1000 Series Oscilloscopes User’s Guide 23The oscilloscope is configured to the following default control settings:Table 2 Au

Página 76

1Getting Started24 Agilent 1000 Series Oscilloscopes User’s GuideStep 6. Compensate probesCompensate probes to match your probe to the input channel.

Página 77

Getting Started 1Agilent 1000 Series Oscilloscopes User’s Guide 25High Frequency CompensationFor the supplied passive probes:1 Using the BNC adapter,

Página 78 - To record waveforms

1Getting Started26 Agilent 1000 Series Oscilloscopes User’s GuideStep 7. Become familiar with the Front Panel ControlsBefore using the oscilloscope,

Página 79 - To play-back waveforms

Getting Started 1Agilent 1000 Series Oscilloscopes User’s Guide 27The definitions of the front panel knobs, keys, and softkeys are as follows:Front P

Página 80 - To select the end frame

1Getting Started28 Agilent 1000 Series Oscilloscopes User’s GuideUsing the Oscilloscope Softkey MenusWhen one of the oscilloscope front panel keys tu

Página 81 - To store recorded waveforms

Getting Started 1Agilent 1000 Series Oscilloscopes User’s Guide 29The [Menu On/Off] front panel key turns off the menu or turns on the last accessed

Página 82 - To import/export recordings

Agilent 1000 Series Oscilloscopes User’s Guide 3 Agilent 1000 Series Oscilloscopes—At a GlanceThe Agilent 1000 Series oscilloscopes are low-cost porta

Página 83 - Trigger level knob

1Getting Started30 Agilent 1000 Series Oscilloscopes User’s GuideStep 8. Become familiar with the oscilloscope displayChannel 1 Status,DC coupling, 5

Página 84 - To force a trigger

Getting Started 1Agilent 1000 Series Oscilloscopes User’s Guide 31Step 9. Use the Run Control keysThere are two front panel keys for starting and sto

Página 85 - To set up edge triggers

1Getting Started32 Agilent 1000 Series Oscilloscopes User’s GuideStep 10. Access the built-in helpThe oscilloscope has built- in quick help informati

Página 86

Getting Started 1Agilent 1000 Series Oscilloscopes User’s Guide 33Securing the OscilloscopeTo secure a 1000 Series oscilloscope to its location, you

Página 87 - To set up video triggers

1Getting Started34 Agilent 1000 Series Oscilloscopes User’s Guide

Página 88

35Agilent 1000 Series OscilloscopesUser’s Guide 2Displaying DataUsing the Horizontal Controls 36Using the Vertical Controls 43Using Math Function W

Página 89 - Capturing Data 3

2 Displaying Data36 Agilent 1000 Series Oscilloscopes User’s GuideUsing the Horizontal ControlsThe horizontal controls consist of:• The horizontal sc

Página 90 - To set up pattern triggers

Displaying Data 2Agilent 1000 Series Oscilloscopes User’s Guide 37To adjust the horizontal scale• Turn the horizontal scale knob to change the horizo

Página 91 - To set up alternate triggers

2 Displaying Data38 Agilent 1000 Series Oscilloscopes User’s GuideWhen the time/div is set to 50 ms/div or slower, the oscilloscope enters Slow Scan

Página 92 - To set the trigger coupling

Displaying Data 2Agilent 1000 Series Oscilloscopes User’s Guide 39• Push the horizontal position knob to “zero” the trigger point (in other words, mo

Página 93

4 Agilent 1000 Series Oscilloscopes User’s Guide In This BookThis guide shows how to use the Agilent 1000 Series oscilloscopes.1Getting StartedDescrib

Página 94

2 Displaying Data40 Agilent 1000 Series Oscilloscopes User’s GuideTo change the horizontal time base (Y-T, X-Y, or Roll)1 Press [Menu/Zoom].2 In the

Página 95 - To specify a trigger holdoff

Displaying Data 2Agilent 1000 Series Oscilloscopes User’s Guide 41X-Y FormatThis format compares the voltage level of two waveforms point by point. I

Página 96 - 3Capturing Data

2 Displaying Data42 Agilent 1000 Series Oscilloscopes User’s GuideThe following modes or functions are not available in X-Y format:• Automatic voltag

Página 97 - Making Measurements

Displaying Data 2Agilent 1000 Series Oscilloscopes User’s Guide 43Using the Vertical ControlsThe vertical controls consist of:• The channel ([1], [2

Página 98

2 Displaying Data44 Agilent 1000 Series Oscilloscopes User’s GuideTo adjust the vertical scaleWhen an input channel waveform is on:• Turn its vertica

Página 99

Displaying Data 2Agilent 1000 Series Oscilloscopes User’s Guide 45Notice that, as you adjust the vertical position, a message showing the position of

Página 100 - Vmax (Maximum Voltage)

2 Displaying Data46 Agilent 1000 Series Oscilloscopes User’s GuideFigure 17 DC Coupling ControlFigure 18 AC Coupling ControlDC CouplingDC Coupling St

Página 101

Displaying Data 2Agilent 1000 Series Oscilloscopes User’s Guide 47To specify a bandwidth limitWhen high frequency components of a waveform are not im

Página 102 - Preshoot

2 Displaying Data48 Agilent 1000 Series Oscilloscopes User’s GuideTo specify the probe attenuationFor correct measurements, you must match the oscill

Página 103 - Time Measurements

Displaying Data 2Agilent 1000 Series Oscilloscopes User’s Guide 49To use a digital filterYou can apply a digital filter to the sampled waveform data.

Página 104 - Rise Time

Agilent 1000 Series Oscilloscopes User’s Guide 5 ContentsAgilent 1000 Series Oscilloscopes—At a Glance 3In This Book 4Figures 13Tab les 151 Getting St

Página 105 - Negative Duty Cycle

2 Displaying Data50 Agilent 1000 Series Oscilloscopes User’s GuideTo change the Volts/Div control sensitivityWhen you need to adjust the amplitude/di

Página 106 - Delay Between Falling Edges

Displaying Data 2Agilent 1000 Series Oscilloscopes User’s Guide 51Figure 21 Waveform Before InversionFigure 22 Waveform After Inversion

Página 107 - Phase Between Falling Edges

2 Displaying Data52 Agilent 1000 Series Oscilloscopes User’s GuideTo specify the channel units1 If the channel’s menu is not currently displayed, pre

Página 108 - Counter (Frequency)

Displaying Data 2Agilent 1000 Series Oscilloscopes User’s Guide 53Using Math Function WaveformsThe math functions control allows the selection of the

Página 109 - Making Cursor Measurements

2 Displaying Data54 Agilent 1000 Series Oscilloscopes User’s GuideTo add, subtract, or multiply waveforms1 Press [Math].2 In the Math menu, press Op

Página 110

Displaying Data 2Agilent 1000 Series Oscilloscopes User’s Guide 555 Press Window, and continue pressing the softkey or turn the entry knob to sele

Página 111

2 Displaying Data56 Agilent 1000 Series Oscilloscopes User’s Guide10 Use the horizontal position knob to adjust the frequency per division.The freque

Página 112

Displaying Data 2Agilent 1000 Series Oscilloscopes User’s Guide 57Using Reference WaveformsYou can save a reference waveform to an internal, nonvolat

Página 113

2 Displaying Data58 Agilent 1000 Series Oscilloscopes User’s Guide3 Presh Location to choose “External”.4 Press Save or Import.5 Use the disk manag

Página 114 - Saving and Recalling Data

Displaying Data 2Agilent 1000 Series Oscilloscopes User’s Guide 59Changing the Display SettingsTo display waveforms as vectors or dots1 Press [Displ

Página 115

6 Agilent 1000 Series Oscilloscopes User’s Guide ContentsSecuring the Oscilloscope 332 Displaying Data 35Using the Horizontal Controls 36To adjust the

Página 116

2 Displaying Data60 Agilent 1000 Series Oscilloscopes User’s GuideTo clear the display1 Press [Display].2 In the Display menu, press Clear.To set wa

Página 117

Displaying Data 2Agilent 1000 Series Oscilloscopes User’s Guide 61To display graded waveform intensityWhile the oscilloscope is running, waveforms sh

Página 118 - Using the Disk Manager

2 Displaying Data62 Agilent 1000 Series Oscilloscopes User’s GuideTo change the menu display timeThe menu display time is how long menus remain on th

Página 119 - To create new folders

Displaying Data 2Agilent 1000 Series Oscilloscopes User’s Guide 63To select screen persistenceThe screen persistence setting specifies what is displa

Página 120 - To edit folder/file names

2 Displaying Data64 Agilent 1000 Series Oscilloscopes User’s Guide

Página 121 - To delete files

65Agilent 1000 Series OscilloscopesUser’s Guide 3Capturing DataOverview of Sampling 66Choosing the Acquisition Mode 73Choosing the Acquisition Mode

Página 122 - To display disk information

3Capturing Data66 Agilent 1000 Series Oscilloscopes User’s GuideOverview of SamplingTo understand the oscilloscope’s sampling and acquisition modes,

Página 123 - USB Host Port

Capturing Data 3Agilent 1000 Series Oscilloscopes User’s Guide 67Oscilloscope Bandwidth and Sample RateAn oscilloscope’s bandwidth is typically descr

Página 124

3Capturing Data68 Agilent 1000 Series Oscilloscopes User’s GuideHowever, digital signals have frequency components above the fundamental frequency (s

Página 125

Capturing Data 3Agilent 1000 Series Oscilloscopes User’s Guide 69So, in practice, an oscilloscope’s sample rate should be four or more times its band

Página 126 - 1 Press [Print]

ContentsAgilent 1000 Series Oscilloscopes User’s Guide 7 Changing the Display Settings 59To display waveforms as vectors or dots 59To clear the displa

Página 127 - Oscilloscope Utility Settings

3Capturing Data70 Agilent 1000 Series Oscilloscopes User’s GuideAn oscilloscope’s rise time is not the fastest edge speed that the oscilloscope can a

Página 128 - Turning Sound ON or OFF

Capturing Data 3Agilent 1000 Series Oscilloscopes User’s Guide 71Real-Time SamplingThe 1000 Series oscilloscopes provide real- time sampling. In othe

Página 129

3Capturing Data72 Agilent 1000 Series Oscilloscopes User’s GuideMemory Depth and Sample RateThe number of points of oscilloscope memory is fixed (exc

Página 130

Capturing Data 3Agilent 1000 Series Oscilloscopes User’s Guide 73Choosing the Acquisition ModeThe osilloscope can operate in normal, average, or peak

Página 131 - To enable/disable mask tests

3Capturing Data74 Agilent 1000 Series Oscilloscopes User’s GuideTo select the Average acquisition modeIn the Average acquisition mode, acquisitions a

Página 132 - To run/stop a mask test

Capturing Data 3Agilent 1000 Series Oscilloscopes User’s Guide 75The Average acquisition mode decreases the screen refresh rate.To select the Average

Página 133 - To set up masks

3Capturing Data76 Agilent 1000 Series Oscilloscopes User’s GuideHowever, in the Peak Detect acquisition mode, acquisitions are made at the fastest sa

Página 134

Capturing Data 3Agilent 1000 Series Oscilloscopes User’s Guide 77To turn OFF/ON sine(x)/x interpolationWhen sample points are displayed as vectors (i

Página 135 - To export/import masks

3Capturing Data78 Agilent 1000 Series Oscilloscopes User’s GuideRecording/Playing-back WaveformsYou can record waveforms from input channels or from

Página 136 - To set up the screen saver

Capturing Data 3Agilent 1000 Series Oscilloscopes User’s Guide 79To start/stop recording1 In the Sequence menu ([Acquire]>Sequence>Mode=Rec

Página 137

8 Agilent 1000 Series Oscilloscopes User’s Guide ContentsAdjusting the Trigger Level 83To adjust the trigger level 83To force a trigger 84Choosing the

Página 138 - Running Self-Calibration

3Capturing Data80 Agilent 1000 Series Oscilloscopes User’s GuideTo select continuous or one-time play-back1 In the Sequence menu ([Acquire]>Sequ

Página 139

Capturing Data 3Agilent 1000 Series Oscilloscopes User’s Guide 81To store recorded waveformsTo store recorded waveforms:1 Press [Acquire].2 In the A

Página 140 - Pollution Degree Definitions

3Capturing Data82 Agilent 1000 Series Oscilloscopes User’s GuideTo load a recording1 In the Sequence menu ([Acquire]>Sequence>Mode=Storage)

Página 141 - Measurement Category

Capturing Data 3Agilent 1000 Series Oscilloscopes User’s Guide 83Adjusting the Trigger LevelTo adjust the trigger level• Turn the trigger [Level] kno

Página 142 - Specifications

3Capturing Data84 Agilent 1000 Series Oscilloscopes User’s GuideTo force a triggerTo make an acquisition even if no valid trigger has been found:1 Pr

Página 143 - Characteristics

Capturing Data 3Agilent 1000 Series Oscilloscopes User’s Guide 85Choosing the Trigger ModeThe trigger determines when captured data should be stored

Página 144

3Capturing Data86 Agilent 1000 Series Oscilloscopes User’s Guide5 Press Source and continue pressing the softkey or turn the entry knob to select

Página 145

Capturing Data 3Agilent 1000 Series Oscilloscopes User’s Guide 875 Press Source and continue pressing the softkey or turn the entry knob to select

Página 146

3Capturing Data88 Agilent 1000 Series Oscilloscopes User’s Guide5 Press Polarity to toggle between:6 Press Sync and continue pressing the softkey or

Página 147 - Table 11 Measurement Features

Capturing Data 3Agilent 1000 Series Oscilloscopes User’s Guide 89Figure 35 Line SynchronizationFigure 36 Field Synchronization

Página 148 - Table 14 I/O

ContentsAgilent 1000 Series Oscilloscopes User’s Guide 9 Voltage Measurements 100Vmax (Maximum Voltage) 100Vmin (Minimum Voltage) 101Vpp (Peak-to-Peak

Página 149 - Table 16 Power Requirements

3Capturing Data90 Agilent 1000 Series Oscilloscopes User’s GuideTo set up pattern triggers1 Press [Menu].2 In the Trigger menu, press Mode.3 Continu

Página 150

Capturing Data 3Agilent 1000 Series Oscilloscopes User’s Guide 91To set up alternate triggersThe Alternate trigger mode splits the display horizontal

Página 151 - Warnings

3Capturing Data92 Agilent 1000 Series Oscilloscopes User’s GuideSetting Other Trigger ParametersThese are trigger system parameters that apply in all

Página 152 - Safety Symbols

Capturing Data 3Agilent 1000 Series Oscilloscopes User’s Guide 934 Continue pressing the the Coupling softkey or turn the entry knob to select one

Página 153

3Capturing Data94 Agilent 1000 Series Oscilloscopes User’s GuideTo change the trigger sensitivityTrigger sensitivity specifies the vertical change th

Página 154

Capturing Data 3Agilent 1000 Series Oscilloscopes User’s Guide 95To specify a trigger holdoffTrigger holdoff can be used to stabilize a waveform. The

Página 155

3Capturing Data96 Agilent 1000 Series Oscilloscopes User’s GuideUsing the External Trigger InputYou can trigger on external inputs by selecting “EXT”

Página 156

97Agilent 1000 Series OscilloscopesUser’s Guide 4Making MeasurementsDisplaying Automatic Measurements 98Voltage Measurements 100Time Measurements 1

Página 157

4 Making Measurements98 Agilent 1000 Series Oscilloscopes User’s GuideDisplaying Automatic MeasurementsYou can use the [Measure] key to display autom

Página 158 - *54130-97014*

Making Measurements 4Agilent 1000 Series Oscilloscopes User’s Guide 99A maximum of three measurements can be displayed at the bottom of the display.