GPIB Command Dictionary
CDANalyzer subsystem
:IS2000:PNUMber:MODE <string>
:IS2000:PNUMber:MODE?
:IS2000:PNumber:MODE <string>
:IS2000:PNumber:MODE?
IS-2000 only. These commands set/query the PN setup.
‘Auto’ is used to calculate the system’s PN offset if it is unknown.
When ‘Auto’ is selected, the CDAN:IS2000:PNUM:INCR command
can be used to enter a PN increment and speed the PN offset
calculation.
‘Manual’ is used to enter a known PN offset. When ‘Manual’ is
selected, the PN offset is entered using the CDAN:IS2000:PN:OFFS
command.
Syntax
CDAN:IS2000:PNUM:MODE ‘Auto’
‘Manual’
CDAN:IS2000:PNUM:MODE? !returns the PN Offset mode
CDAN:IS2000:PN:MODE ‘Auto’
‘Manual’
CDAN:IS2000:PN:MODE? !returns the PN Offset mode
Screen/field equivalent
CDAN:IS2000:PNUM:MODE and CDAN:IS2000:PN:MODE control the
Find PN field on the PN Setup image of the CODE DOMAIN screen
when the CDMA standard is set to “IS-2000”.