Chapter 1 251
GPIB Command Dictionary
MEASure subsystem
:CDANalyzer:IS2000:TRACe:POWer[:BASic]:ACTive?
IS-2000 only. This query returns the active channel information for the
Code Domain trace when making a Power, Fast Power or Fast Power
Synchronize measurement. The return entity consists of an integer
representing the number of active channels followed by a four value
record for each channel. The four values for each record are channel
number, spread factor for this channel, Power for this channel and a
fourth value that should be ignored.
Syntax
MEAS:CDAN:IS2000:TRAC:POW[:BAS]:ACT?
Screen/field equivalent
There is no equivalent field for this query on the CODE DOMAIN
screen. The information is accessible only when a Power, Fast Power,
orPower & Noise measurement is being displayed, CD pwr unit is set
to Rel from the Reference image on the CODE DOMAIN screen and
the CDMA standard is set to “IS-2000”.