Chapter 1 155
GPIB Command Dictionary
CONFigure subsystem
:OFLevel:RFINout <real number>
:OFLevel:RFINout?
These commands set/query the path loss from the RF IN/OUT
connector to the device-under-test. This correction is applied when the
CONF:OFL:MODE ‘On’ command is used.
This command utilizes the “Real Number Setting Syntax” on page 299,
but does not use the :STATe command.
Syntax
CONF:OFL:RFIN <real number> !values: minus 100.0 to 100.0
CONF:OFL:RFIN?
Screen/field equivalent
CONF:OFL:RFIN controls the RF In/Out field of the INSTRUMENT
CONFIGURE screen. This value in this field is used when the RF
Level Offset field is set to On.
:OFRequency <real number>
:OFRequency?
These commands set/query the frequency offset between the Test Set’s
generator and analyzer. This command is used only in frequency tuning
mode (CONF:RFD ‘Freq’) and RF offset must be turned on
(CONF:OMODE ‘On’).
This command utilizes the “Real Number Setting Syntax” on page 299,
but does not use the :STATe command.
Syntax
CONF:OFR <real number>
CONF:OFR?
Screen/field equivalent
CONF:OFR controls the (Gen)-(Anl) field of the INSTRUMENT
CONFIGURE screen.