304 Chapter 1
GPIB Command Dictionary
RFANalyzer subsystem
:FREQuency <real number>
:FREQuency?
These commands set/query the tune frequency for the RF analyzer. This
command requires that the CONF:RFD ‘Freq’ command is used.
This command utilizes the “Real Number Setting Syntax” on page 299,
but does not use the :STATe command.
Syntax
RFAN:FREQ <real>
RFAN:FREQ?
Screen/field equivalent
RFAN:FREQ corresponds to the Tune Freq field on the RF ANALYZER
screen.
[:FREQuency]:GTIMe <real number>
[:FREQuency]:GTIMe?
These commands set/query the gate time for the RF Frequency counter.
This command utilizes the :DUNits and UNITs commands of the “Real
Number Setting Syntax” on page 299.
Syntax
RFAN:FREQ:GTIM <real>
RFAN:FREQ:GTIM?
Screen/field equivalent
RFAN:FREQ:GTIM controls the RF Cnt Gate field on the RF
ANALYZER screen.