106 Chapter 1
GPIB Command Dictionary
CANalyzer subsystem
:PATH:GAIN <string>
:PATH:GAIN?
These commands set/query the gain of the CDMA analyzer. This
command will only be enabled when autoranging is turned off
(CAN:PATH:MODE ‘Hold’).
This command may conflict with the code domain analyzers settings
CDAN:PATH:GAIN and CDAN:PATH:MODE. For example, if
autoranging is turned off (‘Hold’) in the CDMA analyzer and turned on
in the Code Domain Analyzer (‘Auto’) the ‘Hold’ setting will always
override the ‘Auto’ setting. When autoranging is turned off, setting
either analyzers gain will automatically change the others to the same
setting.
Use the CAN:POW:GAIN command when making an average power
measurement.
Syntax
CAN:PATH:GAIN ‘0 dB’
‘6 dB’
‘12 dB’
‘18 dB’
‘24 dB’
‘30 dB’
‘36 dB’
CAN:PATH:GAIN?
Screen/field equivalent
CAN:PATH:GAIN sets the lower subfield of the Gain field of the CDMA
ANALYZER screen. This subfield is only selectable when the upper
subfield is set to Hold.