290 Chapter 1
GPIB Command Dictionary
OSCilloscope subsystem
:SCALe:VERTical:OFFSet <real number>
:SCALe:VERTical:OFFSet?
These commands set/query the vertical scale (DC) offset of the vertical
axis of the oscilloscope display. This moves the signal up to four
divisions up or down with respect to the oscilloscope’s fixed center line.
This command utilizes the “Real Number Setting Syntax” on page 299,
but does not use the :STATe command.
Syntax
OSC:SCAL:VERT:OFFS <real number>
OSC:SCAL:VERT:OFFS?
Example
OSC:SCAL:VERT:OFFS 2.5 !moves the signal 2.5 divisions
Screen/field equivalent
OSC:SCAL:VERT:OFFS controls the Vert Offset field on the Main
menu of the SCOPE screen.