Chapter 1 243
GPIB Command Dictionary
MEASure subsystem
:CDANalyzer:IS2000:TRACe:ABSolute:POWer[:BASic]:ALL?
IS-2000 only. This query returns the information necessary to recreate
the Code Domain trace when making a Power, Fast Power, or Fast
Power Synchronize measurement. The returned entity consists of 128
four-value records for each bar in the trace. The four values for each
record are:
Channel number. If the selected location is part of a supplemental
channel, all the Walsh channels that comprise the supplemental
channel will be returned.
Power for this channel. If this is a supplemental channel, the total
power for the supplemental channel will be returned.
A third (ignore it) value
A flag indicating that the next bar should be connected to this bar as
part of a wider supplemental channel.
Syntax
MEAS:CDAN:IS2000:TRAC:ABS:POW[:BAS]:ALL?
Screen/field equivalent
MEAS:CDAN:IS2000:TRAC:ABS:POW[:BAS]:ALL? has no equivalent
field on the CODE DOMAIN screen. The information can only be
returned when the CDMA standard is set to “IS-2000”, the
measurement selected is Power, Fast Power or FstPwr Sync, and CD
pwr unit is set to Abs on the Reference image of the CODE DOMAIN
screen.
Vista de página 242