Agilent Technologies 8935 Series Manual do Utilizador Página 201

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 352
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 200
Chapter 1 201
GPIB Command Dictionary
DECoder subsystem
:DTMF:GATE <real number>
:DTMF:GATE?
This command specifies how long the decoder analyzes a signal after it
has been triggered. Up to 6.55 seconds of gate time can be specified.
If the gate time is too long, the decoders buffer becomes full. A message
is displayed instructing you to decrease the gate time.
This command utilizes the “Real Number Setting Syntax” on page 299,
but does not use the :STATe command.
Syntax
DEC:DTMF:GATE <real number>
DEC:DTMF:GATE?
Screen/field equivalent
DEC:DTMF:GATE controls the Gate Time field on the SIGNALING
DECODER screen when the Mode field is set to DTMF.
:FGEN:GATE <real number>
This command specifies how long the decoder analyzes a signal after it
has been triggered. Up to 6.55 seconds of gate time can be specified.
If the gate time is too long, the decoders buffer becomes full. A message
is displayed instructing you to decrease the gate time.
This command utilizes the “Real Number Setting Syntax” on page 299,
but does not use the :STATe command.
Syntax
DEC:FGEN:GATE <real number>
Screen/field equivalent
DEC:FGEN:GATE controls the Gate Time field on the SIGNALING
DECODER screen when the Mode field is set to Func Gen.
Vista de página 200
1 2 ... 196 197 198 199 200 201 202 203 204 205 206 ... 351 352

Comentários a estes Manuais

Sem comentários