1 Power Meter Remote Operation
70 N1911A/1912A P-Series Power Meters Programming Guide
Using Device Clear to Halt Measurements
Device clear is an IEEE- 488 low- level bus message which can be used to
halt measurements in progress. Different programming languages and
IEEE- 488 interface cards provide access to this capability through their
own unique commands. The status registers, the error queue, and all
configuration states are left unchanged when a device clear message is
received. Device clear performs the following actions.
All measurements in progress are aborted.
The power meter returns to the trigger “idle state”.
The power meter’s input and output buffers are cleared.
The power meter is prepared to accept a new command string.
NOTE
For interfaces the that do not support a low-level device clear, use the ABORt command.