Agilent Technologies 1912A Manual de Serviço Página 236

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 803
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 235
4 CALibration Subsystem
208 N1911A/1912A P-Series Power Meters Programming Guide
P- Series power sensor is connected the error –241, “Hardware missing”
occurs.
If the calibration was not carried out successfully the error –231, “Data
Questionable; CAL ERROR” occurs. If you are using an N1912A the
error message specifies which channel failed calibration.
If there is no sensor connected, the error –241, “Hardware Missing”
occurs.
If this command is set to ON and TRIGger[SEQuence[1]|2]:COUNt
is set to a value >1, the error –221, “Setting conflict” occurs.
If this command is set to ON (for dual channel, at either measurement
channel) when a P- Series power sensor is connected (for dual channel,
at either measurement channel) and is in the wait- for- trigger state for
external trigger buffering, the error –224, “Illegal parameter value”
occurs.
Vista de página 235
1 2 ... 231 232 233 234 235 236 237 238 239 240 241 ... 802 803

Comentários a estes Manuais

Sem comentários