10 SENSe Subsystem
396 N1911A/1912A P-Series Power Meters Programming Guide
Query
[SENSe[1]]|SENSe2:CORRection:CSET[1]|CSET2:STATe?
The query returns a 1 or 0 into the output buffer indicating whether a
table is enabled or disabled.
1 is returned when the table is enabled
0 is returned when the table is disabled
Query Example
Error Messages
If you attempt to set this command to ON and no table has been
selected using
[SENSe[1]]|SENSe2:CORRection:CSET[1]|CSET2:[SELect] then error
–221, “Settings conflict” occurs and
[SENSe[1]]|SENSe2:CORRection:CSET[1]|CSET2:STATe remains OFF.
If the CSET1 command is used when a P- Series, N8480 Series
(excluding Option CFT) or an E-Series power sensor is connected, the
error –241 “Hardware missing” occurs.
SENS2:CORR:CSET1:STAT? This command queries whether there is
currently an active sensor calibration
table for Channel B.