CALCulate Subsystem 3
N1911A/1912A P-Series Power Meters Programming Guide 193
CALCulate[1]|2|3|4:PHOLd:CLEar
This command clears the peak hold value for a specified CALC block so
that a new peak hold value can be set.
Syntax
Example
Error Messages
If no power sensor is connected, error –241 “Hardware missing” occurs.
If a sensor, other than an P- Series or E9320 power sensor, is
connected, error –241 “Hardware missing” occurs.
•If SENS:DET:FUNC is set to AVER or TRIG:SOUR is set to INT1, INT2 or
EXT, error –221 “Settings conflict” occurs.
NOTE
Clearing the peak hold value for a specified CALC block may affect the peak hold value of
other CALC blocks, depending on the CALC channel set up (set by CALC:MATH:EXPR).
CALC2:PHOLd:CLEar This command clears the peak hold value
for CALC2.
:PHOL
:CLE
CALC
1
2
3
4