9 PSTatistic Subsystem
324 N1911A/1912A P-Series Power Meters Programming Guide
Query
PSTatistic:CCDF:REFerence[:STATe]?
The query enters a 1 or 0 into the output buffer indicating the status of
the reference trace stored.
1 is returned when the previously stored reference trace is turned on
0 is returned when the previously stored reference trace is turned off
Query Example
Error Messages
If command is executed in other window besides expanded CCDF
window, error –221 "Settings conflict: Requires CCDF expanded
window" occurs.
If there was no previously saved trace, error –221 "Settings conflict: No
reference trace saved" occurs. Please check the status of saved reference
trace using command PST:CCDF:STOR:REF?.
PST:CCDF:REF? This command queries whether the
reference trace is turned on or off.