Agilent Technologies 1912A Manual de Serviço Página 128

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 803
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 127
2 MEASurement Commands
100 N1911A/1912A P-Series Power Meters Programming Guide
CONFigure[1]|2|3|4[:SCALar][:POWer:AC]:RELative
[<expected_value>[,<resolution>[,<source list>]]]
This command sets the measurement function, range and resolution of the
specified window. It sets the measurement function to single channel with
relative mode on. The relative value used is that set by the
CALCulate:RELative:MAGNitude:AUTO command.
Syntax
Parameters
Refer to “Optional Parameters” on page 90 for additional details on the
parameters in this command.
:POW
:AC
:SCAL
:REL
Space
expected_value
DEF
resolution
,
DEF
source list
,
CONF
1
2
3
4
Item Description/Default Range of Values
expected_value
(for the expected
power level)
A numeric value for the expected power
level. The units of measurement are dBm
and W. The default units are defined by
UNIT:POWer.
sensor dependent
DEF
1
resolution A numeric value for the resolution. If
unspecified the current resolution setting
is used.
1 to 4
2
1.0, 0.1, 0.01, 0.001
DEF
1
Vista de página 127
1 2 ... 123 124 125 126 127 128 129 130 131 132 133 ... 802 803

Comentários a estes Manuais

Sem comentários