Agilent Technologies 1912A Manual de Serviço Página 416

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 803
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 415
10 SENSe Subsystem
388 N1911A/1912A P-Series Power Meters Programming Guide
[SENSe[1]]|SENSe2:CORRection:CFACtor|GAIN[1][:INPut][:M
AGNitude] <numeric_value>
This command is used to enter a gain correction value for the calibration
factor. The power meter corrects every measurement by this factor to
compensate for the gain.
Either CFACtor and GAIN1 can be used in the commandboth have an
identical result. Using GAIN1 complies with the SCPI standard, whereas
CFACtor does not—this may make your program easier to understand.
Syntax
SENS
1
CORR
SENS2
:
:INP
:MAGN
Space
?
numeric_value
DEF
MIN
MAX
MIN
MAX
Space
:GAIN
1
:CFAC
Vista de página 415
1 2 ... 411 412 413 414 415 416 417 418 419 420 421 ... 802 803

Comentários a estes Manuais

Sem comentários