PSTatistic Subsystem 9
N1911A/1912A P-Series Power Meters Programming Guide 341
Example
Error Messages
If no power sensor is connected, error –241 "Hardware missing" occurs.
If sensor/sensors connected are not P- Series sensors, error –241
"Hardware missing" occurs.
If measurement speed setting is FAST, error –221 "Settings conflict"
occurs.
If the acquisition mode is in continuous triggering or triggering source
is set to either INT1, INT2, or EXT, error –221 "Setting conflict"
occurs.
If the parameter specified is less than 0.0 or more than 50.0, error –220
"Parameter error" occurs.
If no parameter is specified, error –109 "Missing parameter" occurs.
PST1:CCDF:PROB? 50 This command queries the probability at