Agilent Technologies 1912A Manual de Serviço Página 467

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 803
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 466
SENSe Subsystem 10
N1911A/1912A P-Series Power Meters Programming Guide 439
Example
Reset Condition
On reset, the value is set to 0 seconds.
Query
SENSe[1]]|SENSe2:SWEep[1]|2|3|4:OFFSet:TIME?
The query returns the current delay between the trigger point and the
start of the time- gated period.
Query Example
Error Messages
If the command is used when a sensor other than a P- Series or E9320
power sensor is connected, error –241, “Hardware missing” occurs
If the command is used when an E9320 sensor is connected and set to
AVERage mode rather than NORMal mode, the error –221, “Settings
Conflict” occurs.
SENS2:SWE3:OFFS:TIME 0.001 This command sets the delay to 0.001
seconds.
SENS2:SWE2:OFFS:TIME? The query returns the current delay
between the trigger point and the start of
the time- gated period for Channel B and
gate 2.
Vista de página 466
1 2 ... 462 463 464 465 466 467 468 469 470 471 472 ... 802 803

Comentários a estes Manuais

Sem comentários