Agilent Technologies 1912A Manual de Serviço Página 241

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 803
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 240
CALibration Subsystem 4
N1911A/1912A P-Series Power Meters Programming Guide 213
CALibration[1]|2:ZERO:AUTO [ONCE|ON|OFF|0|1]
This command causes the power meter to perform its zeroing routine on
the specified channel when enabled. This adjusts the power meter for a
zero power reading with no power supplied to the power sensor.
1|ON can only be used with a P-Series sensor. When 1|ON is enabled the
the zero is maintained by a combination of on- the- fly zero measurements
and temperature compensation.
The 0|OFF parameter is only required for the query response and is
ignored in the command.
Except when using a P- Series sensor, this command assumes that a power
sensor is not connected to a power source.
Syntax
Example
Reset Condition
On reset, automatic zeroing is disabled.
CAL2:ZERO:AUTO ONCE This command causes the power meter to
perform a zeroing routine on Channel B.
:AUTO
:ZERO
0|OFF
1|OFF
Space
?
CAL
1
2
ONCE
Vista de página 240
1 2 ... 236 237 238 239 240 241 242 243 244 245 246 ... 802 803

Comentários a estes Manuais

Sem comentários