Agilent Technologies E5071C Guia do Utilizador Página 62

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 551
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 61
62 Chapter 4
Performing a Calibration
Calibration
Calculating the Calibration Coefficients
To calculate the calibration coefficients, use the following command:
Before issuing the above command, you must measure all the required calibration data
items according to your selected calibration types (see Table 4-1).
Calculating the calibration coefficients clears all calibration data regardless it is used for
the calculation. The calibration type selection is also cleared, which results in the state
when no calibration type is selected.
Turning On or Off Error Correction
To turn on or off Error Correction, use the following command:
:SENS{1-4}:CORR:STAT on page 412
Also, once you have calculated the calibration coefficient using the
:SENS{1-4}:CORR:COLL:SAVE command, Error Correction is automatically turned
on.
Calibration type Command
All calibration types :SENS{1-4}:CORR:COLL:SAVE on page 407
Vista de página 61
1 2 ... 57 58 59 60 61 62 63 64 65 66 67 ... 550 551

Comentários a estes Manuais

Vikiehs 04 Jan 2024 | 13:20:06

urenrjrjkvnm