Chapter 13 425
SCPI Command Reference
E5061A/E5062A commands
13. SCPI Command
Reference
:SENS{1-4}:SWE:POIN
Syntax :SENSe{[1]|2|3|4}:SWEep:POINts <numeric>
:SENSe{[1]|2|3|4}:SWEep:POINts?
Description Sets the number of measurement points of channel 1 (:SENS1) to channel 4 (:SENS4).
Parameters
If the specified parameter is out of the allowable setup range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response {numeric}<newline><^END>
Example of use
10 OUTPUT 717;":SENS1:SWE:POIN 801"
20 OUTPUT 717;":SENS1:SWE:POIN?"
30 ENTER 717;A
Equivalent key [Sweep Setup] - Points
<numeric>
Description Number of measurement points
Range 2 to 1601
Preset value 201
Resolution 1