Chapter 13 435
SCPI Command Reference
E5061A/E5062A commands
13. SCPI Command
Reference
:SOUR{1-4}:POW:PORT{1-2}
Syntax :SOURce{[1]|2|3|4}:POWer:PORT{[1]|2}[:LEVel][:IMMediate][:AMPLitude] <numeric>
:SOURce{[1]|2|3|4}:POWer:PORT{[1]|2}[:LEVel][:IMMediate][:AMPLitude]?
Description Sets the power level of port 1 (:PORT1) to port 2 (:PORT2) of channel 1 (:SOUR1) to
channel 4 (:SOUR4).
Parameters
If the specified parameter is out of the allowable setup range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response {numeric}<newline><^END>
Example of use
10 OUTPUT 717;":SOUR1:POW:PORT1 -12.5"
20 OUTPUT 717;":SOUR1:POW:PORT1?"
30 ENTER 717;A
Related commands :SOUR{1-4}:POW:PORT:COUP on page 434
:SOUR{1-4}:POW:ATT on page 432
Equivalent key
[Sweep Setup] - Power - Port Power - Port 1 Power|Port 2 Power
<numeric>
Description The power level at the specified port.
Range Varies depending on the power range.
Preset value 0
Unit dBm
Resolution 0.05