Chapter 13 415
SCPI Command Reference
E5061A/E5062A commands
13. SCPI Command
Reference
:SENS{1-4}:FREQ:CENT
Syntax :SENSe{[1]|2|3|4}:FREQuency:CENTer <numeric>
:SENSe{[1]|2|3|4}:FREQuency:CENTer?
Description Sets the center value of the sweep range of channel 1 (:SENS1) to channel 4 (:SENS4).
Parameters
If the specified parameter is out of the allowable setup range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response {numeric}<newline><^END>
Example of use
10 OUTPUT 717;":SENS1:FREQ:CENT 2E9"
20 OUTPUT 717;":SENS1:FREQ:CENT?"
30 ENTER 717;A
Related commands :SENS{1-4}:FREQ:SPAN on page 417
Equivalent key
[Center]
<numeric>
Description Center value
Range 3E5 to 3E9
Preset value 4.25015E9
Unit Hz (hertz)
Resolution 0.5 or 1