406 Chapter 13
SCPI Command Reference
E5061A/E5062A commands
:SENS{1-4}:CORR:COLL:OPEN
Syntax :SENSe{[1]|2|3|4}:CORRection:COLLect[:ACQuire]:OPEN <numeric>
Description For channel 1 (:SENS1) to channel 4 (:SENS4), measures the calibration data of the open
standard of the specified port. (No query)
Parameters
Example of use
10 OUTPUT 717;":SENS1:CORR:COLL:OPEN 1"
20 OUTPUT 717;"*OPC?"
30 ENTER 717;A
Equivalent key [Cal] - Calibrate - Response (Open)|1-Port Cal - Open
[Cal] - Calibrate - n-Port Cal - Reflection - Port m Open
<numeric>
Description Port number
Range 1 to 2