Chapter 13 243
SCPI Command Reference
E5061A/E5062A commands
13. SCPI Command
Reference
:CALC{1-4}:MARK:BWID
Syntax :CALCulate{[1]|2|3|4}[:SELected]:MARKer:BWIDth[:STATe] {ON|OFF|1|0}
:CALCulate{[1]|2|3|4}[:SELected]:MARKer:BWIDth[:STATe]?
Description For the active trace of channel 1 (:CALC1) to channel 4 (:CALC4) (specified with the
:CALC{1-4}:PAR{1-4}:SEL command), turns ON/OFF the bandwidth search result
display.
Parameters
Query response {1|0}<newline><^END>
Example of use
10 OUTPUT 717;":CALC1:MARK:BWID ON"
20 OUTPUT 717;":CALC1:MARK:BWID?"
30 ENTER 717;A
Related commands :CALC{1-4}:MARK{1-10}:BWID:DATA? on page 265
:CALC{1-4}:MARK{1-10}:BWID:THR on page 266
:CALC{1-4}:PAR{1-4}:SEL on page 286
Equivalent key
[Marker Search] - Bandwidth
Description
ON or 1 Turns ON the bandwidth search result display.
OFF or 0 (preset value) Turns OFF the bandwidth search result display.