Chapter 3 53
Setting Up the Analyzer
Sample Program
3. Setting Up the Analyzer
Channel 2 Sweep type Linear
Sweep range Center value 1.9 GHz
Span value 500 MHz
Number of measurement points 101
IF bandwidth 30 kHz
Power 0 dBm
Number of traces 4
Graph Layout Two graphs at left and right
Trace 1 Measurement parameter S21
Data format Log magnitude
Reference division line
number
9
Reference division line
value
2
Scale per division 10 dBm
Trace 2 Measurement parameter S22
Data format Smith chart (Lin)
Full-scale value 2
Table 3-2 Segment table for channel 1
Segment
Number
Start value Stop value
Number of
measurement
points
IF bandwidth Power
1 1.7 GHz 1.9 GHz 21 30 kHz 0 dBm
2 1.9 GHz 2 GHz 101 10 kHz -10 dBm
3 2 GHz 2.2 GHz 21 30 kHz 0 dBm
Table 3-1 Target settings in Example 3-1
Item Setting