Agilent Technologies E5071C Guia do Utilizador Página 402

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 551
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 401
402 Chapter 13
SCPI Command Reference
E5061A/E5062A commands
:SENS{1-4}:CORR:COLL:METH:OPEN
Syntax :SENSe{[1]|2|3|4}:CORRection:COLLect:METHod[:RESPonse]:OPEN <numeric>
Description For channel 1 (:SENS1) to channel 4 (:SENS4), sets the calibration type to the response
calibration (open) of the specified port. (No query)
Parameters
Example of use
10 OUTPUT 717;":SENS1:CORR:COLL:METH:OPEN 1"
Related commands :SENS{1-4}:CORR:COLL:METH:TYPE? on page 405
Equivalent key
[Cal] - Calibrate - Response (Open) - Port
:SENS{1-4}:CORR:COLL:METH:SHOR
Syntax :SENSe{[1]|2|3|4}:CORRection:COLLect:METHod[:RESPonse]:SHORt <numeric>
Description For channel 1 (:SENS1) to channel 4 (:SENS4), sets the calibration type to the response
calibration (short) of the specified port. (No query)
Parameters
Example of use
10 OUTPUT 717;":SENS1:CORR:COLL:METH:SHOR 1"
Related commands :SENS{1-4}:CORR:COLL:METH:TYPE? on page 405
Equivalent key
[Cal] - Calibrate - Response (Short) - Port
<numeric>
Description Port number
Range 1 to 2
Resolution 1
<numeric>
Description Port number
Range 1 to 2
Resolution 1
Vista de página 401
1 2 ... 397 398 399 400 401 402 403 404 405 406 407 ... 550 551

Comentários a estes Manuais

Vikiehs 04 Jan 2024 | 13:20:06

urenrjrjkvnm