Agilent Technologies E5071C Guia do Utilizador Página 379

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 551
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 378
Chapter 13 379
SCPI Command Reference
E5061A/E5062A commands
13. SCPI Command
Reference
:SENS{1-4}:CORR:COLL:CKIT:ORD:SHOR
Syntax :SENSe{[1]|2|3|4}:CORRection:COLLect:CKIT:ORDer:SHORt <numeric 1>,<numeric
2>
:SENSe{[1]|2|3|4}:CORRection:COLLect:CKIT:ORDer:SHORt? <numeric 1>
Description For the calibration kit selected for channel 1 (:SENS1) to channel 4 (:SENS4), selects the
standard used for the short measurement of the specified port.
Parameters
If the specified parameter is out of the allowable setup range, an error occurs and the
command is ignored.
Query response {numeric 2}<newline><^END>
Example of use
10 OUTPUT 717;":SENS1:CORR:COLL:CKIT:ORD:SHOR 1,1"
20 OUTPUT 717;":SENS1:CORR:COLL:CKIT:ORD:SHOR? 1"
30 ENTER 717;A
Related commands :SENS{1-4}:CORR:COLL:CKIT on page 375
Equivalent key
[Cal] - Modify Cal Kit - Specify CLSs - Short - Port 1|Port 2
<numeric 1> <numeric 2>
Description Port number Standard number
Range 1 to 2 1 to 21
Resolution 1 1
Vista de página 378
1 2 ... 374 375 376 377 378 379 380 381 382 383 384 ... 550 551

Comentários a estes Manuais

Vikiehs 04 Jan 2024 | 13:20:06

urenrjrjkvnm