378 Chapter 13
SCPI Command Reference
E5061A/E5062A commands
:SENS{1-4}:CORR:COLL:CKIT:ORD:OPEN
Syntax :SENSe{[1]|2|3|4}:CORRection:COLLect:CKIT:ORDer:OPEN <numeric 1>,<numeric 2>
:SENSe{[1]|2|3|4}:CORRection:COLLect:CKIT:ORDer:OPEN? <numeric 1>
Description For the calibration kit selected for channel 1 (:SENS1) to channel 4 (:SENS4), selects the
standard used for the open measurement of the specified port.
Parameters
If the specified parameter is out of the allowable setup range, an error occurs and the
command is ignored.
Query response {numeric 2}<newline><^END>
Example of use
10 OUTPUT 717;":SENS1:CORR:COLL:CKIT:ORD:OPEN 1,2"
20 OUTPUT 717;":SENS1:CORR:COLL:CKIT:ORD:OPEN? 1"
30 ENTER 717;A
Related commands :SENS{1-4}:CORR:COLL:CKIT on page 375
Equivalent key
[Cal] - Modify Cal Kit - Specify CLSs - Open - Port 1|Port 2
<numeric 1> <numeric 2>
Description Port number Standard number
Range 1 to 2 1 to 21
Resolution 1 1