414 Chapter 13
SCPI Command Reference
E5061A/E5062A commands
:SENS{1-4}:FREQ
Syntax :SENSe{[1]|2|3|4}:FREQuency[:CW|FIXed] <numeric>
:SENSe{[1]|2|3|4}:FREQuency[:CW|FIXed]?
Description Sets the fixed frequency (CW frequency) for the power sweep for channel 1 (:SENS1) to
channel 4 (:SENS4).
Parameters
If the specified parameter is out of the allowable setup range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response {numeric}<newline><^END>
Example of use
10 OUTPUT 717;":SENS1:FREQ 1E9"
20 OUTPUT 717;":SENS1:FREQ?"
30 ENTER 717;A
Related commands :SENS{1-4}:SWE:TYPE on page 428
Equivalent key
[Sweep Setup] - Power - CW Freq
<numeric>
Description Fixed frequency
Range 3E5 to 3E9
Preset value 3E5
Unit Hz (hertz)
Resolution 1