Agilent Technologies E5071C Guia do Utilizador Página 398

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 551
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 397
398 Chapter 13
SCPI Command Reference
E5061A/E5062A commands
:SENS{1-4}:CORR:COLL:ECAL:THRU
Syntax :SENSe{[1]|2|3|4}:CORRection:COLLect:ECAL:THRU <numeric 1>,<numeric 2>
Description Executes response calibration (thru) between the 2 specified ports of channel 1 (:SENS1)
to channel 4 (:SENS4) using the ECal (Electrical Calibration) module.
If you execute this command when the ECal module is not connected, an error occurs and
the command is ignored. (No query)
Parameters
For each parameter, you must specify a different port number. If you specify the same port
number for 2 or more parameters, an error occurs and the command is ignored.
Example of use
10 OUTPUT 717;":SENS1:CORR:COLL:ECAL:SOLT2 1,2"
20 OUTPUT 717;"*OPC?"
30 ENTER 717;A
Equivalent key [Cal] - ECal - Thru Cal - 2-1 (S21)|1-2 (S12)
<numeric 1> <numeric 2>
Description Response port number Stimulus port number
Range 1 to 2 1 to 2
Resolution 1 1
Vista de página 397
1 2 ... 393 394 395 396 397 398 399 400 401 402 403 ... 550 551

Comentários a estes Manuais

Vikiehs 04 Jan 2024 | 13:20:06

urenrjrjkvnm