Agilent Technologies E5071C Guia do Utilizador Página 290

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 551
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 289
290 Chapter 13
SCPI Command Reference
E5061A/E5062A commands
:CALC{1-4}:RLIM:DISP:VAL
Syntax :CALCulate{[1]|2|3|4}[:SELected]:RLIMit:DISPlay:VALue {OFF|ABS|MAR}
:CALCulate{[1]|2|3|4}[:SELected]:RLIMit:DISPlay:VALue?
Description For the active trace of channel 1 (:CALC1) to channel 4 (:CALC4) (specified with the
:CALC{1-4}:PAR{1-4}:SEL command), selects the display type for the ripple value.
Parameters
Query response {OFF|ABS|MAR}<newline><^END>
Example of use
10 OUTPUT 717;":CALC1:RLIM:DISP:VAL ABS"
20 OUTPUT 717;":CALC1:RLIM:DISP:VAL?"
30 ENTER 717;A$
Related commands :CALC{1-4}:PAR{1-4}:SEL on page 286
:CALC{1-4}:RLIM on page 286
:CALC{1-4}:RLIM:DISP:LINE on page 288
:CALC{1-4}:RLIM:DISP:SEL on page 289
Equivalent key
[Analysis] - Ripple Limit - Ripple Value - OFF | Absolute | Margin
Description
OFF (preset value) Specifies the display off.
ABS Specifies the absolute value for display type.
MAR Specifies the margin for display type.