Repair and Maintenance
Recovering from an Operating System Crash
Figure 40 Add Bootp Entry
5 Exit with OK.
6 Select Configure -> Bootp Manager. All entries made above are shown in
Figure 41.
Figure 41 Bootp Manager
7 Press Apply to activate the changes. The Bootp Manager now shows the
entries made above.
8 Press OK to exit the Bootp Manager and power cycle the VW detector, to force
it to send a new bootp request again. This time, the MAC address will be
recognized by the Bootp Server (Figure 42). It will send the configured IP
address and subnet mask information which are necessary for
communication to the JetDirect card.