Agilent Technologies 1100 Series Manual do Utilizador Página 43

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 42
43
Setting up an Analysis
Requirements and Conditions
Requirements and Conditions
What You Will Need
Table 6 lists the items you need to have for the set up of the analysis. Some of
these are optional (not required for the basic system).
Conditions
A single injection of the isocratic test standard is made under the conditions
given in Table 7:
Table 6 What you will need
HP 1100 Instrument Isocratic pump
Manual Injector
Detector - VWD, Standard flow cell installed
ChemStation or Control Module (optional)
Column: Zorbax Eclipse XDB-C8 150 x 4.6 mm i.d.
Part No. 993967-902 or Part No. 5063-6600
Standard: Part No. 01080-68704
0.15 wt.% dimethylphthalate, 0.15 wt.% diethylphthalate,
0.01 wt.% biphenyl, 0.03 wt.% o-terphenyl in methanol
Table 7 Conditions
Flow: 1.5 ml/minute
Stoptime: 6 minutes
Solvent: 100% (30% water/70% Acetonitrile)
Temperature Ambient
Wavelength VWD: 254 nm
Injection Volume 2 µl (VWD)
Vista de página 42
1 2 ... 38 39 40 41 42 43 44 45 46 47 48 ... 75 76

Comentários a estes Manuais

Sem comentários