Agilent Technologies 1100 Series Manual do Utilizador Página 13

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 12
13
General information about your system
Site Requirements
Bench Space
The modules dimensions and weight (see Table 2 on page 14) allow to place
the modules on almost any desk or laboratory bench. It needs an additional
2.5 cm (1.0 inches) of space on either side and approximately 8 cm
(3.1 inches) in the rear for the circulation of air and electric connections.
If the bench should carry a complete system, make sure that the bench is
designed to carry the weight of all the modules.
The modules should be operated in a horizontal position.
Environment
Your modules will work within specifications at ambient temperatures and
relative humidity as described in Table 2 on page 14.
Detectors
ASTM drift tests require a temperature change below 2 °C/hour (3.6 °F/hour)
over one hour period. Our published drift specification (refer also to
Performance Specifications in your detectors reference manuals) is based
on these conditions. Larger ambient temperature changes will result in larger
drift. Better drift performance depends on better control of the temperature
fluctuations. To realize the highest performance, minimize the frequency and
the amplitude of the temperature changes to below 1 °C/hour (1.8 °F/hour).
Turbulences around one minute or less can be ignored.
CAUTION Do not store, ship or use your modules under conditions where temperature
fluctuations could cause condensation within the modules. Condensation will
damage the system electronics. If your modules were shipped in cold weather,
leave them in their boxes and allow them to warm slowly to room temperature
to avoid condensation.