Installing your System
Choose the optimal stack configuration
Choose the optimal stack configuration
Figure 2 You can ensure optimum performance by installing the system in the
following configuration. This configuration optimizes the system flow path
and ensures minimum delay volume.
NOTE An additional thermostatted column compartment (G1316A) must be located
between the pump and the detector.
An additional detector (G1321A FLD or G1362A RID) must be located under
the VW detector to be the last in the flow path (pressure limit on those flow
cells).
An additional local control module (G1323B) may be added to one of the
modules.
Solvent container
Pump
VW detector
Manual injector
Vista de página 19