Agilent Technologies 53230A Especificações Página 96

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 95
Chapter 2 Service
About the Agilent 53210A/53220A/53230A Calibration Menu
96 Assembly-Level Service Guide
About the Agilent 53210A/53220A/53230A
Calibration Menu
NOTE If you are familiar with the information provided in this section,
proceed to the section titled “The Agilent 53210A/53220A/53230A
Calibration Procedures” on page 88 to calibrate your counter.
This section, “About the Agilent 53210A/53220A/53230A
Calibration Menu,” can be used as a reference. Your counter
should be calibrated at least once each year to maintain
specification accuracy.
Overview of the Agilent 53210A/53220A/53230A
Calibration Menu
The Calibration menu is accessed by pressing the Utility key and then the
Cal / AutoCal softkey. All of the calibration factors and security settings
are stored in non-volatile memory, and do not change when power has
been off or after a remote interface reset.
Vista de página 95
1 2 ... 91 92 93 94 95 96 97 98 99 100 101 ... 259 260

Comentários a estes Manuais

Sem comentários