Chapter 1 Performance Tests
Agilent 53210A/53220A/53230A Complete Performance Tests
Assembly-Level Service Guide 61
4Press the Gate key.
•The Gate softkeys are displayed:
5 Press and set:
Gate Src: Timed
Gate Out: On
•Gate Time: 10 mSec
6Press the Trigger key.
•The Trigger softkeys are displayed:
7 Press to set:
Source: Manual
Wait For Trig is displayed
Rdgs/Trigger: 100
Trig Delay: 0 seconds