Chapter 1 Performance Tests
Agilent 53210A/53220A/53230A Complete Performance Tests
Assembly-Level Service Guide 75
14 Press to set:
• Statistics: On
• The display shows:
Channel 3 Pulse Burst Frequency Test Procedure
1Press the Trigger key.
• The counter takes 100 readings and fills in the math functions.
2 Verify the ‘Mean’ is:
• 1.35 GHz ± 1.35 MHz
3 Verify the ‘Std Dev’ is less than:
• 338 kHz