Chapter 6 Backdating
Backdating Firmware
Assembly-Level Service Guide 251
Figure 6-1. Information Screen
NOTE If any number other than 01.07 or 01.08 is displayed as the number
before the first dash of the Firmware version, then refer to
“Procedure Changes” for information on calibration and/or
performance test procedures that apply to your counter.
NOTE If Option 106 “Microwave Input” is installed in your counter and
the “FW Rev” is (27.0), refer to “Procedure Changes” for
information concerning the calibration procedure for your
counter.