Chapter 3 Replacing Assemblies
To Remove the GPIB Assembly
Assembly-Level Service Guide 177
To Remove the GPIB Assembly
1 Remove the cover and rear bezel from the counter as described
previously.
2 If installed, remove the Option 300 Internal DC Battery assembly
as described previously.
3 Carefully remove the ribbon cable from J1 on the GPIB printed
circuit board.
4 Using a 7mm spintite, remove the two hex screws securing the
GPIB assembly to the rear panel.
5 Remove the GPIB assembly from the counter.
6 To replace the GPIB assembly, reverse the above procedure.
Figure 3-4. Top View of Counter showing GPIB assembly