Chapter 2 Service
The Agilent 53210A/53220A/53230A Calibration Procedures
118 Assembly-Level Service Guide
Figure 2-6. DC Voltage Source and DMM Connections
11 Press the Perform Cal Step softkey.
If the calibration step PASSES, after a few seconds the counter
momentarily displays:
“Calibration step succeeded.”
–The Cal Count on the screen will increment.
If the calibration step FAILS, after a few seconds the counter displays:
“Computed calibration correction factor out of range” and also
“Calibration step failed.
After several seconds, “Calibration step failed” is cleared but
“Computed calibration correction factor out of range” remains.
–The Cal Count does not update.
Press the Back - Local key to clear the “Computed calibration
correction factor out of range” message.
NOTE If the calibration fails, the current calibration constants remain
unchanged.
DC
Voltage
Source
DMM
BNC
“ T ”