Chapter 2 Service
The Agilent 53210A/53220A/53230A Calibration Procedures
118 Assembly-Level Service Guide
Figure 2-6. DC Voltage Source and DMM Connections
11 Press the Perform Cal Step softkey.
• If the calibration step PASSES, after a few seconds the counter
momentarily displays:
– “Calibration step succeeded.”
–The Cal Count on the screen will increment.
• If the calibration step FAILS, after a few seconds the counter displays:
– “Computed calibration correction factor out of range” and also
“Calibration step failed”.
– After several seconds, “Calibration step failed” is cleared but
“Computed calibration correction factor out of range” remains.
–The Cal Count does not update.
• Press the Back - Local key to clear the “Computed calibration
correction factor out of range” message.
NOTE If the calibration fails, the current calibration constants remain
unchanged.
DC
Voltage
Source
DMM
BNC
“ T ”