!"Agilent N9340B Handheld Spectrum AnalyzerUser’s Guide

Página 2 -

1Overview4 N9340B User’s GuideOptimized UsabilityThe Agilent N9340B handheld spectrum analyzer provides enhanced usability:• The 6.5-inch TFT colorfu

Página 3 - Contents

5Key Reference94 N9340B User’s GuideCF StepKey access: [FREQ] > {CF Step}Changes the step size for the center frequency function. Once a step size

Página 4 - 5 Key Reference

Key Reference 5N9340B User’s Guide 95MarkerAccesses the marker control softkeys to select the type and number of markers. Markers are diamond- shaped

Página 5 - 8 Menu Map

5Key Reference96 N9340B User’s GuideDeltaKey access: [MARKER] > {Delta}Activates a second marker at the position of the first marker. (If no marker

Página 6 - Documentation Conventions:

Key Reference 5N9340B User’s Guide 97Peak SearchPlace a marker on the highest peak based on the setting of the Search Criteria as follows:•PeakKey acc

Página 7

5Key Reference98 N9340B User’s GuideMarker ToAccesses the following marker function menu keys:•To CenterKey access: [MARKER] > {To Center}Sets the

Página 8 - Introduction

Key Reference 5N9340B User’s Guide 99Marker TraceKey access: [MARKER] > {More} > {Marker Trace}Activates a marker on the trace if there are no m

Página 9

5Key Reference100 N9340B User’s GuideMEASHardkey menu access.In spectrum analysis mode, pressing [MEAS] brings up the menu for a suite of power measu

Página 10 - Optimized Usability

Key Reference 5N9340B User’s Guide 101ACPRThe adjacent channel power ratio (ACPR) measures the power ratio between the main channel power and the adja

Página 11 - Front Panel Overview

5Key Reference102 N9340B User’s GuideChannel PowerChannel power function measures the power and power spectral density in the channel bandwidth specif

Página 12 - 15:44:03 03,10,08

Key Reference 5N9340B User’s Guide 103Spectrum Emission Mask (SEM)Spectrum Emission Mask measurement includes the in- band and out- of- band spurious

Página 13 - Top Panel Overview

Overview 1N9340B User’s Guide 5Front Panel OverviewN9340B HANDHELD SPECTRUM ANALYZER 100 kHz - 3.0 GHz SYS MODE MEAS TRACEBW/SWPAMPTD SPAN FREQEN

Página 14 - Instrument Markings

5Key Reference104 N9340B User’s Guide• Chan SpanKey access: [MEAS] > {SEM} > {Ref Channel}Specifies the span used in measuring the power in the

Página 15 - Getting Started

Key Reference 5N9340B User’s Guide 105Table 2 Offsets & Limits Defaulted for W-CDMA (3GPP) Mobile Station Measurementsaa. Abs Stop is set to Coupl

Página 16

5Key Reference106 N9340B User’s GuideOffsetKey access: [MEAS] > {SEM} > {Offset/Limits}Selects the offset pairs (upper and lower) that the softk

Página 17 - Power Requirements

Key Reference 5N9340B User’s Guide 107Specifies the resolution bandwidth used in measuring the offset pair. When set to Auto, the Res BW from the defa

Página 18 - AC Power Cords

5Key Reference108 N9340B User’s GuideMeas BWKey access: [MEAS] > {SEM} > {Offset/Limits} > {More} > {More}Specifies the bandwidth to use w

Página 19 - Safety Considerations

Key Reference 5N9340B User’s Guide 109SpectrogramSpectrogram Run/PauseKey access: [MEAS] > {Spectrogram} > {Spectrogram}Toggles the spectrogram

Página 20 - 14 N9340B User’s Guide

5Key Reference110 N9340B User’s GuidePaletteKey access: [MEAS]> {Spectrogram} > {Palette}Toggles the spectrogram color mode between full color a

Página 21 - Electrical Requirements

Key Reference 5N9340B User’s Guide 111File SetupKey access: [MEAS]> {Spectrogram} > {Save&Load} > {File Setup}This submenu allows you to

Página 22 - Working with Batteries

5Key Reference112 N9340B User’s GuideMODEActivates the softkey menus to select the measurement mode of your analyzer. The measurement modes are:• Spec

Página 23 - Charging a Battery

Key Reference 5N9340B User’s Guide 113Amptd StepKey access: [MODE] > {Tracking Generator } > { Amptd Step}Allows you to set the step size of the

Página 24 - Press Power Switch

1Overview6 N9340B User’s GuideDisplay Annotations123456789101211141315:44:03 03,10,08Description Associated Function Key1 Time and Date [SYS] > {Se

Página 25 - Preparation for Use

5Key Reference114 N9340B User’s Guidethe error message: Store reference trace before turning on Normalize (On) is displayed in the status line, howeve

Página 26 - Entering Data

Key Reference 5N9340B User’s Guide 115such as an amplifier. Now replace the thru line with the device under test, and an accurate measurement of the g

Página 27

5Key Reference116 N9340B User’s Guide• Zero TypeToggles the zeroing type of the Power Sensor between internal and external. When this type is set to E

Página 28 - 22 N9340B User’s Guide

Key Reference 5N9340B User’s Guide 117•ResolutionAccess the associated Resolution submenu.When the measurement suffix is W or %, this parameter repres

Página 29 - System and File

5Key Reference118 N9340B User’s GuideMeas SetupKey access: [MODE] > {Power Meter} > {Meas Setup}Accesses the Meas Disp submenu which is used to

Página 30 - System Setting

Key Reference 5N9340B User’s Guide 119• Lower LimitSets the lower test limit value for the current measurement. The units used are dependent on the cu

Página 31 - Factory Default Settings

5Key Reference120 N9340B User’s Guideaverages is greater than 12.5%, the digital filter is cleared. The filter will then start storing new measurement

Página 32 - General system settings

Key Reference 5N9340B User’s Guide 121Quick SpanKey access: [MODE] > {IBOC Measurement} > {AM}/{FM} > {Quick Span}Press this softkey to quick

Página 33 - IP configuration

5Key Reference122 N9340B User’s GuideMask FuctionKey access: [MODE] > {IBOC Measurement} > {AM}/{FM} > {Mask Function}Accesses the submenu to

Página 34 - Ext Input

Key Reference 5N9340B User’s Guide 123The mak types for IBOC- FM are listed as below: • FCC (FM- 73.317)• NRSC- 5- B Hybrid• NRSC- 5- B HybridFigure 5

Página 35 - Adding an Option

Overview 1N9340B User’s Guide 7Top Panel OverviewPC12-18 VDC 55W M AX 50 VDC MA X 33 dBm (2 W) MAXRF IN PUT 50 RF OU T 50 EXT TRIG IN/EXT

Página 36 - Show Error

5Key Reference124 N9340B User’s Guide• BeepBeep On/Off is a switch to enable/disable Pass/Fail beep sound indication.•Noise CancelThis function is use

Página 37 - Low Frequency Channel

Key Reference 5N9340B User’s Guide 125Tune RangeKey access: [MODE] > {IBOC Measurement} > {AM}/{FM} > {Tune Range}Toggles Tune Range between

Página 38 - Viewing file list

5Key Reference126 N9340B User’s Guidespectrogram. when the Spectrogram is running, Pressing {Spec View}will pause the running spectrogram. Press {Ho

Página 39 - Saving a file

Key Reference 5N9340B User’s Guide 127Timed SettingKey access: [MEAS]> {Spectrogram} > {Save&Load} > {Time Setting}This submenu allows yo

Página 40 - Quick saving a setup

5Key Reference128 N9340B User’s GuideSPANThis hardkey accesses the softkey submenu of span functions. Pressing [SPAN] allows you to change the frequen

Página 41 - Loading a file

Key Reference 5N9340B User’s Guide 129AMKey access: [SPAN] > {Demod} > {AM}Pressing {AM ON} activates the AM demodulation and turns off FM demo

Página 42 - 36 N9340B User’s Guide

5Key Reference130 N9340B User’s GuideHz. For long demodulation times, pressing Preset when Preset Type is set to DFT will abort the Demod function. T

Página 43 - Making Measurements

Key Reference 5N9340B User’s Guide 131Ext RiseKey access: [SPAN] > {Trigger} > {Ext Rise}Activates the trigger condition that will start a sweep

Página 44 - Measuring Multiple Signals

5Key Reference132 N9340B User’s GuideTRACEAccesses the trace keys to store and manipulate trace information. Each trace comprises a series of data poi

Página 45 - Accuracy" on page 48

Key Reference 5N9340B User’s Guide 133Minimum HoldKey access: [TRACE] > {Min Hold}Maintains the minimum level for each trace point of the selected

Página 46 - Signal generator

1Overview8 N9340B User’s GuideInstrument MarkingsThe CE mark shows that the product complies with all relevant European Legal Directives.The CSA mar

Página 47

5Key Reference134 N9340B User’s Guideof the data evaluated during that time and present a single point of trace data based on the detector mode. We ca

Página 48 - Hidden by Large Signals

Key Reference 5N9340B User’s Guide 135AverageKey access: [TRACE] > {More} > {Average}Initiates a digital averaging routine that averages the tra

Página 49 - Measuring a Low-Level Signal

5Key Reference136 N9340B User’s GuideNumbers, alpha characters and underline are all valid components of a filename.Recall TraceKey access: [TRACE] &g

Página 50

Key Reference 5N9340B User’s Guide 137LIMITLimit LineKey access: [Limit] > {Limit Line}Activates an adjustable horizontal line that is used as a vi

Página 51

5Key Reference138 N9340B User’s GuideLimit TypeKey access: [Limit] > {Limit Type}Defines the limit you are editing as either an upper or lower limi

Página 52 - Increased Sweep Time

139Agilent N9340B Handheld Spectrum AnalyzerAgilent Technologies6Error MessagesThis

Página 53 - Trace Averagi ng

6 Error Messages140 N9340B User’s GuideOverviewIf an improper operation occurs during the instrument configuration, a message will appear on the statu

Página 54

Error Messages 6N9340B User’s Guide 141Error Message ListPlease refer to the error messages list table as below to find the corresponding message desc

Página 55

6 Error Messages142 N9340B User’s Guide–144 Character data too longIndicates the character data contained more than 12 characters. (see IEEE 488.2, 7.

Página 56

Error Messages 6N9340B User’s Guide 143150 Mixer overloadIndicates the first mixer in danger. Either increase input attenuation or decrease the input

Página 57

9Agilent N9340B Handheld Spectrum AnalyzerAgilent Technologies2Getting Star

Página 58

6 Error Messages144 N9340B User’s Guide

Página 59

145Agilent N9340B Handheld Spectrum AnalyzerAgilent Technologies7TroubleshootingThi

Página 60

7 Troubleshooting146 N9340B User’s GuideCheck the basicsThis section includes information on how to check for a problem with your Agilent spectrum ana

Página 61 - CAU-CAUTION

Troubleshooting 7N9340B User’s Guide 147• To meet specifications, the analyzer must be aligned. For more information on how to align the analyzer, ref

Página 62

7 Troubleshooting148 N9340B User’s GuideContact Agilent TechnologiesAgilent has Sales and Service Offices around the world to provide you with complet

Página 63 - Low-pass Filter

149Agilent N9340B Handheld Spectrum Analyzer Agilent Technologies8Menu MapThis chapt

Página 64

8Menu Map150 N9340B User’s GuideAMPTDAMPTDRef levelAttenuationAuto ManPreampOff OnScale/DIV >Scale TypeLog LinHiSensitivityMoredBm1 dB/DIV2 d

Página 65 - N9340A

BW/SWPRBWVBWVBW/RBWAvg TypeSweep TimeAuto ManSweepSingle SweepLog PwrPowerVoltageTD ModeAuto ManAuto ManAuto ManSweep Single ContSweep ModeN

Página 66

FreqCenter FreqCF StepAuto ManStart FreqStop FreqChannel STD>StandardOff On Load STD ><Chan name>ChannelFwd RevChannel StepSTD T

Página 67 - N9340B User’s Guide 61

MarkerMarkerNormalDeltaOffPeak Search >Marker To >MorePeak 6 Peaks Search1 2 3 4 5 6Next Left PKNext Right PK1 of 2To C en te rTo R efMode >

Página 68 - Calibration Routine

2 Getting Started10 N9340B User’s GuideChecking Shipment and Order ListCheck the shipment and order list when you receive the shipment.• Inspect the s

Página 69

8Menu Map154 N9340B User’s GuideMEAS MEASMeasure OffACPRChannel PowerFieldStrengthOBWMethod%dBcCenter FreqInt BWChannel BWCenter FreqAdj ChannelChanne

Página 70

Menu Map 8N9340B User’s Guide 155MODE - IBOC MeasurementIBOC Measur ementFM>Qu ick S pan >Auto TuneReturnCenter Freq<val ue >Mask Functio

Página 71

8Menu Map156 N9340B User’s GuideMODE - Tracking GeneratorMODESpectrum AnalyzerSourcePower MeterTracking GeneraotrDemodulation AnalysisAmplitudeOff

Página 72

Menu Map 8N9340B User’s Guide 157MODE - Power Meter Zeroing Meas Disp Re tu rn Meas Setup Preset Freq Power Range Return Limits Rel/Offset Avera

Página 73

8Menu Map158 N9340B User’s GuideMODE - Demodulation Analysis (AM/FM)Demodula tion AnalysisIF BWDe tec torMoreCarrier FreqEq LP FX Scale*Ref ValueRetur

Página 74

Menu Map 8N9340B User’s Guide 159MODE - Demodulation Analysis (ASK/FSK)Demodulation AnalysisSymbol RateViewMoreCarrier FreqFilter SetupX Scale*Ref Val

Página 75

8Menu Map160 N9340B User’s GuideMODE - xDSLMODESpectrogramRun PauseRestart MeasOff OnSpectrg ViewHomePage UpPage DownEndPlayback xDSLPalette Dis

Página 76 - Spectrogram Audio Alert

Menu Map 8N9340B User’s Guide 161SPANSPANSpanFull SpanZero SpanLast SpanDemod*Trigger*The trigger menu is available when span is set to zero span.AMOf

Página 77 - Demodulating an FM Signal

8Menu Map162 N9340B User’s GuideSYSSYSBrightness Setting >Key BacklightFile >GeneralInput/OutputTrig RefLanguageMore1 of 3Show SystemShow Error

Página 78

TRACETrac eClear WriteMax HoldMin HoldViewBlankMore1 2 3 4 1 of 2DetectorSave TraceAverageNormalAveragePos PeakMore2 of 2Auto ManOff OnSave AsReca

Página 79 - AM/FM Demodulation Analysis

Getting Started 2N9340B User’s Guide 11Power RequirementsThe AC power supplied must meet the following requirements:The AC/DC power supply charger ada

Página 80 - FM Demodulation Analysis

Limit TypeBeepSave LimitsUpper LowerLIMITLimit LineLimitLoad LimitsOff OnOff OnOff OnLimits PointsLimits TableLimits OffsetLimitLimitEdit Limit

Página 81

165AAdding Option, 29Amplitude, 82Attenuation, 82Avg Type, 87BBattery, 16CCF Step, 94Checking SYS Info, 19DDisplay Adjustment, 26EESC/CLR, 92

Página 83

DECLARATION OF CONFORMITY According to EN ISO/IEC 17050-1:2004 Manufacturer’s Name: Agilent Technologies (Chengdu) Co., Ltd. Manufacturer’s Addre

Página 84 - FSK Demodulation Analysis

!"www.agilent.com© Agilent Technologies, Inc. 2008-2009Printed in China July 2009*N9340-90008*N9340-90008

Página 85

Plug Type Cable Part NumberPlug a DescriptionFor use in Country & Region250V 10A8121-1703 BS 1363/A Option 900United Kingdom, Hong Kong, Singapore

Página 86

Getting Started 2N9340B User’s Guide 13Safety ConsiderationsAgilent has designed and tested the N9340B handheld spectrum analyzer for Measurement, Con

Página 87 - Key Reference

Notices© Agilent Technologies, Inc. 2008-2009No part of this manual may be reproduced in any form or by any means (including electronic storage and re

Página 88 - Amplitude

2 Getting Started14 N9340B User’s GuideWARNINGDanger of explosion if the battery is incorrectly replaced. Replace only with the same type battery reco

Página 89 - Scale Type

Getting Started 2N9340B User’s Guide 15Environmental RequirementsThe N9340B is designed for use under the following conditions:• Operating temperature

Página 90 - Y Axis Units

2 Getting Started16 N9340B User’s GuideWorking with BatteriesInstalling a BatteryViewing Battery StatusDetermine the battery status using either of th

Página 91 - Grid Scale

Getting Started 2N9340B User’s Guide 17Charging a BatteryYou may charge the battery both in the tester and in the external battery charger (option BCG

Página 92

2 Getting Started18 N9340B User’s GuidePowering on the Analyzer for the First TimeInsert the battery into the analyzer or connect the analyzer to an e

Página 93 - Avg Type

Getting Started 2N9340B User’s Guide 19Preparation for UseUse [SYS] hardkey to check or set the system settings of your analyzer. Setting up your N93

Página 94

2 Getting Started20 N9340B User’s GuideMaking a Basic MeasurementThis section provides information on basic analyzer operations. It assumes that you a

Página 95

Getting Started 2N9340B User’s Guide 21Viewing a Signal on the Analyzer1 Use a signal generator to generate a CW signal of 1.0 GHz, at a power level o

Página 96 - Sweep Mode

2 Getting Started22 N9340B User’s Guide

Página 97 - Single Sweep

23Agilent N9340B Handheld Spectrum AnalyzerAgilent Technologies3System and FileThis chapter provides the information on basic system setting and file

Página 98

1OverviewIntroduction 2Front Panel Overview 5Display Annotations 6Top Panel Overview 7Instrument Markings 82 Getting StartedChecking Sh

Página 99 - Frequency

3 System and File24 N9340B User’s GuideSystem SettingThis section provides you the basic system configuration which is frequently used before or after

Página 100 - Freq Offset

System and File 3N9340B User’s Guide 25Factory Default SettingsParameter Default SettingCenter Frequency 1.5 GHzStart Frequency 0.0 HzStop Frequency

Página 101

3 System and File26 N9340B User’s GuideVisual and Audio AdjustmentDisplay AdjustmentPress [SYS] > {Brightness} to toggles the screen brightness bet

Página 102

System and File 3N9340B User’s Guide 27• LCD The analyzer turns off the LCD display after a user- defined idle time. Press any key to re- activate the

Página 103 - Peak Search

3 System and File28 N9340B User’s GuideUpgrading FirmwareFollow the steps below to update the N9340B firmware:1 Download the latest N9340B firmware fr

Página 104 - Marker To

System and File 3N9340B User’s Guide 29External Reference (Ref)Use the external reference function as follows:1 Input a 10 MHz signal to the EXT TRIG

Página 105 - Marker Table

3 System and File30 N9340B User’s GuideShow SystemPressing [SYS] > {More} > {Show system} displays the following hardware, software and battery

Página 106 - Selecting a Method

System and File 3N9340B User’s Guide 31When the calibration function is triggered, the current measurement is interrupted and a gauge displays on the

Página 107 - Center Frequency

3 System and File32 N9340B User’s GuideFilePressing [SYS] > {File} accesses to the menu that allows you to manage the file saving and loading.View

Página 108 - Channel Power

System and File 3N9340B User’s Guide 33• CSV (*.CSV) A CSV file records the trace data and available for review on PC.• Setup (*.SET) A setup file re

Página 109 - Spectrum Emission Mask (SEM)

Low Frequency Channel 31File 324 Making MeasurementsMeasuring Multiple Signals 38Measuring a Low-Level Signal 43Improving Frequency Resolu

Página 110 - Offset/Limits

3 System and File34 N9340B User’s GuideSaving Multiple Test SetupsTest Setups that are unique and are necessary to recall frequently can be saved as a

Página 111

System and File 3N9340B User’s Guide 35Quick loading a setup• Load defaultPress [SYS] > {More} > {More} > {LoadSYS Setup} > {Load default

Página 112 - Sweep Time

3 System and File36 N9340B User’s Guide

Página 113 - Rel StopAmpt

37Agilent N9340B Handheld Spectrum Analyzer!"4Making Measurements

Página 114 - Fail Mask

4 Making Measurements38 N9340B User’s GuideMeasuring Multiple SignalsThis section provides information on measuring multiple signals.Comparing Signals

Página 115 - Spectrogram

Making Measurements 4N9340B User’s Guide 39Search} softkey. In this example the next peak is 100 MHz, a harmonic of the 50 MHz signal:• Press [MARKER

Página 116 - Timed Setting

4 Making Measurements40 N9340B User’s GuideResolving Signals of Equal AmplitudeIn this example a decrease in resolution bandwidth is used in combinati

Página 117 - Load Spctrg

Making Measurements 4N9340B User’s Guide 41As you decrease the resolution bandwidth, you improve the resolution of the individual signals and it also

Página 118 - Trackin g Generator

4 Making Measurements42 N9340B User’s GuideResolving Small Signals Hidden by Large SignalsThis example uses narrow resolution bandwidths to resolve t

Página 119 - Normalize

Making Measurements 4N9340B User’s Guide 43Measuring a Low-Level SignalThis section provides information on measuring low- level signals and distingui

Página 120

LIMIT 1376Error MessagesOverview 140Error Message List 1417 TroubleshootingCheck the basics 146Warranty and Service Options 147Contact

Página 121 - Power Meter

4 Making Measurements44 N9340B User’s Guide5 Reduce the span to 1 MHz and if necessary re- center the peak.• Press [SPAN] > 1 > {MHz}6 Set the

Página 122 - Meas Disp

Making Measurements 4N9340B User’s Guide 45Decreasing the Resolution BandwidthResolution bandwidth settings affect the level of internal noise without

Página 123

4 Making Measurements46 N9340B User’s GuideUsing the Average Detector and Increased Sweep TimeThe analyzer’s noise floor response may mask low- level

Página 124 - Meas Setup

Making Measurements 4N9340B User’s Guide 47Trace Averagi ngAveraging is a digital process in which each trace point is averaged with the previous swee

Página 125 - is applied to the

4 Making Measurements48 N9340B User’s GuideImproving Frequency Resolution and AccuracyUsing the frequency counter to improve frequency resolution and

Página 126 - IBOC Measurement

Making Measurements 4N9340B User’s Guide 49Making Distortion MeasurementsThis section provides information on measuring and identifying signal distort

Página 127 - Quick Span

4 Making Measurements50 N9340B User’s Guide3 Change the center frequency to the value of the second (400 MHz) harmonic:• Press [MARKER] > {Peak Se

Página 128 - Mask Fuction

Making Measurements 4N9340B User’s Guide 51Figure 4-14 Identifying Analyzer Distortion (O dB atten)8 Press [AMPTD] > {Attenuation} > 10 > {d

Página 129

4 Making Measurements52 N9340B User’s GuideThird-Order Intermodulation DistortionTwo- tone, third- order intermodulation (TOI) distortion is a common

Página 130 - Auto Tune

Making Measurements 4N9340B User’s Guide 53After the performance of the source/analyzer combination has been verified, the DUT (device under test, for

Página 131 - Spec View

Documentation Conventions:1. A pair of curly brackets { } indicates a softkey, for example {Start} refers to the Start softkey.2. A pair of square bra

Página 132

4 Making Measurements54 N9340B User’s Guide• Press [MARKER] > {Peak Search} > {Next Left (Right) PK}8 Measure the other distortion product:• Pre

Página 133

Making Measurements 4N9340B User’s Guide 55Making a Stimulus Response Transmission MeasurementThe procedure below describes how to use a built- in tra

Página 134 - Last Span

4 Making Measurements56 N9340B User’s Guide5 Press [BW/SWP] > {Sweep Time (Auto)} to put the sweep time into stimulus response auto coupled mode.

Página 135 - Delay Time

Making Measurements 4N9340B User’s Guide 57Measuring Stop Band Attenuation of a Low-pass FilterWhen measuring filter characteristics, it is useful to

Página 136 - Free Run

4 Making Measurements58 N9340B User’s Guide8 Reconnect the DUT to the analyzer. Note that the units of the reference level have changed to dB, indicat

Página 137 - Trig Delay

Making Measurements 4N9340B User’s Guide 59Making a Reflection Calibration MeasurementThe following procedure makes a reflection measurement using a c

Página 138 - Max Hold

4 Making Measurements60 N9340B User’s Guide2 Connect the tracking generator output of the analyzer to the directional bridge or coupler.3 Connect the

Página 139 - Detector

Making Measurements 4N9340B User’s Guide 61Figure 4-21 Short Circuit Normalized

Página 140

4 Making Measurements62 N9340B User’s GuideMeasuring Return Loss Using the Reflection Calibration RoutineThis procedure uses the reflection calibratio

Página 141 - Save Trace

Making Measurements 4N9340B User’s Guide 63Making an Average Power MeasurementAverage power measurements provide a key metric in transmitter performan

Página 142 - Recall Trace

1Agilent N9340B Handheld Spectrum Analyzer!"1Overview

Página 143 - Edit Limits

4 Making Measurements64 N9340B User’s Guideexternal zeroing for input signals below -30 dBm for best accuracy.Making a Basic Average Power Measuremen

Página 144 - Load Limits

Making Measurements 4N9340B User’s Guide 65To Make a Basic Average Power MeasurementYou can follow the steps below to make a basic average power measu

Página 145 - Error Messages

4 Making Measurements66 N9340B User’s GuideSetting Power Meter ResolutionYou can choose from four levels of Power Meter resolution. Higher resolutions

Página 146 - Overview

Making Measurements 4N9340B User’s Guide 67Setting the Power Meter’s Upper and Lower LimitsThe internal Power Meter can be configured to detect when a

Página 147 - Error Message List

4 Making Measurements68 N9340B User’s GuideSpectrogramThe scrolling three- dimensional display is noted for its ability to track the frequency and pow

Página 148

Making Measurements 4N9340B User’s Guide 69Press [FREQ], {Center Freq}, [1.8052], {GHz}, [SPAN], [5], {MHz}.3 Turn on the Spectrogram function.Press

Página 149

4 Making Measurements70 N9340B User’s Guideoffending event or interference. The spectrogram data may be played back for review on either the analyzer’

Página 150 - 144 N9340B User’s Guide

Making Measurements 4N9340B User’s Guide 71Demodulating an FM SignalThis section demonstrates how to demodulate and listen to an FM signal using the b

Página 151 - Troubleshooting

4 Making Measurements72 N9340B User’s Guide6 Set the resolution bandwidth to capture the full bandwidth of the FM signal. To calculate the required ba

Página 152 - Check the basics

Making Measurements 4N9340B User’s Guide 73Analysis the Demodulated SignalsThe N9340B provides the optional AM/FM (Option AMA) and ASK/FSK (Option- DM

Página 153 - Warranty and Service Options

1Overview2 N9340B User’s GuideIntroductionThe Agilent N9340B is a handheld spectrum analyzer with a frequency range of 100 kHz to 3 GHz, tunable to 9

Página 154 - Contact Agilent Technologies

4 Making Measurements74 N9340B User’s Guide4 Set the attenuation state to Auto (default):Press {More 1 of 3} > {More 2 of 3} > {Attenuation Aut

Página 155 - Menu Map

Making Measurements 4N9340B User’s Guide 753 Press {Carries Freq} > [100] > {MHz} to set the carrier frequency.4 Set the attenuation state to A

Página 156 - 150 N9340B User’s Guide

4 Making Measurements76 N9340B User’s GuideASK/FSK Demodulation AnalysisThe ASK/FSK demodulation analysis function supports four display modes: • Symb

Página 157 - N9340B User’s Guide 151

Making Measurements 4N9340B User’s Guide 773 Press {Carries Freq} > [100] > {MHz} to set the carrier frequency.4 Press {Symbol Rate} > [10]

Página 158 - 152 N9340B User’s Guide

4 Making Measurements78 N9340B User’s GuideFSK Demodulation AnalysisPlease refer to the procedures below to analysis the ASK demodulation.1 Connect th

Página 159 - Peaks Search

Making Measurements 4N9340B User’s Guide 79Monitoring the ADSL Signal with Option XDMxDSL (ADSL and VDSL) is widely used for broadband internet connec

Página 160 - 154 N9340B User’s Guide

4 Making Measurements80 N9340B User’s GuidePlease refer to the procedures below for monitoring the ADSL network.1 Press [SYS] > {Setting} > {Pw

Página 161 - MODE - IBOC Measurement

81Agilent N9340B Handheld Spectrum AnalyzerAgilent Technologies5Key ReferenceThis Ch

Página 162 - MODE - Tracking Generator

5Key Reference82 N9340B User’s GuideAmplitudeKey access: [AMPTD]Activates the reference level function and access the associated softkeys to set funct

Página 163 - MODE - Power Meter

Key Reference 5N9340B User’s Guide 83PreampKey access: [AMPTD] > {Preamp}Toggles the internal preamp between On and Off.Preamp results in a correct

Página 164 - 158 N9340B User’s Guide

Overview 1N9340B User’s Guide 3• IBOC Measurement (Option IBC)Provides the in- band- on- channel measurement capability through a dedicated measuremen

Página 165 - N9340B User’s Guide 159

5Key Reference84 N9340B User’s GuideHiSensitivityKey access: [AMPTD] > {HiSensitivity}Press this softkey On to set 0 dB attenuation, - 50 dBm refer

Página 166 - MODE - xDSL

Key Reference 5N9340B User’s Guide 85EMF ModeKey access: [AMPTD] > {More}EMF mode is available when the Y Axis Unit is dBmV, dBμV or V. This mode i

Página 167 - N9340B User’s Guide 161

5Key Reference86 N9340B User’s GuideBW/SWPActivates the bandwidth function and accesses the associated softkeys to control resolution bandwidth, video

Página 168 - Language>

Key Reference 5N9340B User’s Guide 87As the video bandwidth is decreased, the sweep time is increased to maintain amplitude calibration. VBW/RBWKey ac

Página 169 - N9340B User’s Guide 163

5Key Reference88 N9340B User’s GuidePwrKey access: [BW/SWP] > {Avg Type} > {Power}Converts the scale from logarithmic to linear power units, and

Página 170 - 164 N9340B User’s Guide

Key Reference 5N9340B User’s Guide 89Sweep TimeKey access: [BW/SWP] > {Sweep Time}Selects the sweep time for the analyzer to tune across the displa

Página 171

5Key Reference90 N9340B User’s GuideIn FFT ModeIn FFT mode, the sweep time default is auto- coupled. The submenu, sweep time is invalid in FFT mode.Sw

Página 172

Key Reference 5N9340B User’s Guide 91Fast sweep mode provides a fast measurement function that decreases the sweep time. Using Fast Sweep will decreas

Página 173

5Key Reference92 N9340B User’s GuideEnterHardkey function:• Terminates data entry and enters into the analyzer numerical values keyed from numeric key

Página 174 - *N9340-90008*

Key Reference 5N9340B User’s Guide 93FrequencyActivates the center frequency function, and accesses the menu of frequency functions. The center freque