Chapter 2 Service
Error Messages
166 Assembly-Level Service Guide
Query Errors (-400...)
These errors indicate licensing issues within the counter firmware. Refer
to the User’s Guide for information.
• +600, “Internal licensing error”
• +601, “License file corrupt or empty””
• +602, “No valid licenses found for this instrument”
• +603, “Some licenses could not be installed”
• +604, “License not found”
Calibration Errors (700...)
The following errors indicate failures that may occur during a
calibration and which are usually caused by the operator or by
the external calibration source.
NOTE If unable to determine the cause of a calibration error, run the
counter Self Test to determine the instrument status:
– [Utility] / (Instr Setup) / (Self Test)
• +701, “Calibration error; security defeated”
• +702, “Calibration error; calibration memory is secured”
• +703, “Calibration error; secure code provided was invalid”
• +704, “Calibration error: secure code too long”
• +705, “Calibration error; calibration aborted”
• +706, “Calibration error: provided value out of range”
• +707, “Calibration error: computed correction factor out of range”
• +708, “Calibration error: signal measurement out of range”
• +709, “Calibration error: no calibration for this function”
• +710, “Calibration error: full scale correction out of range”
• +711, “Calibration error: calibration string too long”
• +712, “Calibration failed”
• +720, “Auto-calibration failed; input signal detected”
• +740, “Calibration data lost: secure state”